New MC mirror

M.Punturo

(2)

New MC mirror?

• Scattering in the end mirror of the IMC is the first cause of the ISYS upgrade

– Poor quality of the substrate has been the diagnosis

• The REOSC company selected by D.Enard and Lyon to perform the polishing of the new MC mirror failed and abandoned the polishing

– “Both offers from WP and REOSC are on a “best effort basis”, meaning the requirements are considered as objectives. This was already the case for the past orders for the existing core optics but a quality reasonably close to our requirements was finally obtained. In a similar way, Reosc accepted our requirements as objectives while committing on slightly degraded requirements.”

• We MUST CONFIDE in the properties of the Faraday Isolator to avoid the fringes that hindered the ITF locking

• What to do?

(3)

Current mirror

• Small mirror:

– 8 cm diam, 3 cm thick – Back face crowded

– Only 360g with all the components

• Mirror unbalanced by the components

• Pendulum elastic constant determined by the coil wires instead of the suspension wires (125 m Cu-Be wires)

Problems in controls, handling and radiation pressure (?)

(4)

Small IMC mirror?

• Two blanks are available with the usual MC size

– They should be retired from REOSC and sent to GS- Lumonics

– A 6 month polishing time is foreseen

– If we are lucky we will have a low scattering mirror with all the control and noise problems we have now

• Can we do better?

(5)

Large substrates

• 6 Large Herasil end substrates are available in Lyon

– 350mm diameter, 100 mm thick

– Unusable for Virgo+, because of the low Q of Herasil (Tama)

• Could we polish one with the IMC requirements? (GS- Lumonics)

• Changes

– New Marionetta

• Maybe we loose the translation degree of freedom

– New reference mass

– Retuning of the suspension – Retuning of the controls

• More standard situation

Referenties

GERELATEERDE DOCUMENTEN

Thus, examining the role of regulatory focus in the procedural fairness arena promises not only to inform when and why retaliatory responses to unfair treatment emerge

The questionnaire contained measurement scales (including operators) with the goal of measuring behavior and testing the research hypotheses. Open questions

De lezer wordt uitgedaagd om in hoofdstuk 2 zonder tussenkomst van de corresponderende bolpatronen het verband te zien tussen de codenaam van een voorwerp en

In this paper I have defended a new version of The Mirror Thesis, the view that the morally evil person is the perverse mirror image of the moral saint. I started, in §1, by

Indeed, the ta- ble shows that the class-I tests (SPRT and Gauss-SSP) have about 5% probability of drawing the wrong conclusion, and that some class-II tests (Gauss-CI and

The science of friction typically starts with theempiri- cal rules formulated by Amontons and Coulomb for elastically deforming, dry contacts, i.e., the force of friction is

In het huidige onderzoek wordt de relatie tussen behandelmotivatie, therapeutische alliantie met de mentor en internaliserende problematiek bij jeugdigen van 12 tot en met 18 jaar

11 The thesis will thus present a theoretically informed study in film language, in which I attempt to explore the relationship between the filmmaker and the viewer while focussing