University of Groningen

Non-Interceptive Beam Current and Position Monitors for a Cyclotron Based Proton Therapy

Facility

Srinivasan, Sudharsan

DOI:

10.33612/diss.149817352

IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from

it. Please check the document version below.

Document Version

Publisher's PDF, also known as Version of record

Publication date:

2021

Link to publication in University of Groningen/UMCG research database

Citation for published version (APA):

Srinivasan, S. (2021). Non-Interceptive Beam Current and Position Monitors for a Cyclotron Based Proton

Therapy Facility. University of Groningen. https://doi.org/10.33612/diss.149817352

Copyright

Other than for strictly personal use, it is not permitted to download or to forward/distribute the text or part of it without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license (like Creative Commons).

Take-down policy

If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim.

Downloaded from the University of Groningen/UMCG research database (Pure): http://www.rug.nl/research/portal. For technical reasons the number of authors shown on this cover page is limited to 10 maximum.

(2)

169

Sudharsan Srinivasan (CV)

Date of Birth 25.01.1990

Citizenship

Indian

Professional Experience

2017-present PhD Student (Early Stage researcher) at the Paul Scherrer

Institut (PSI), Villigen, Switzerland

2014-2016

Graduate research assistant at Forschungszentrum Jülich,

Germany

2011-2013

Assistant manager at Coal India Limited, Nagpur, India

Education

2017-2021 External Ph.D. student at the Faculty of Medical Sciences at the

University of Groningen, Netherlands

2013-2016 M.Sc. in Nuclear applications from the University of Applied

Science, Aachen, Germany

2007-2011 B.Tech in Mechanical engineering from the National Institute

of Technology, Raipur, India

Referenties

GERELATEERDE DOCUMENTEN

Zorginstituut Nederland Pakket Datum 27 oktober 2016 Onze referentie ACP 63-2 5 5 aHUS (3).. Meest voorkomende complicaties:

Non-Interceptive Beam Current and Position Monitors for a Cyclotron Based Proton Therapy Facility..

For beam loss measurements at the coupling points of the gantries to the beamline, the halo monitors provide a beam current resolution of 10 pA with the lowest-detection threshold

Network analysis is performed to achieve maximum power transfer from the beam with a parametric investigation of the resonator dimensions such as pickup position, dielectric

The linear drives are used to position the stretched wire at multiple positions in order to measure the position dependence of the beam current monitor response and later

From the above observation, a cavity BPM built as a combination of four floating cavities offers a higher signal level for a given position offset and higher position

In addition, measurements were performed by sweeping the beam current at two different energies: 138 MeV and 200 MeV at a position offset of 4.8 mm to check the influence of

Verder wordt een deel van de onderzoekactiviteit besteed aan complete nok- mechanismen. Hierbij wordt vooral gekeken naar de kwaliteit van de voortge- brachte beweging en de