Three-dimensional quantitative coronary angiography and the registration with intravascular ultrasound and optical

coherence tomography

Tu, S.

Citation

Tu, S. (2012, February 28). Three-dimensional quantitative coronary angiography and the registration with intravascular ultrasound and optical coherence tomography. ASCI dissertation series. Retrieved from

https://hdl.handle.net/1887/18531

Version: Corrected Publisher’s Version

License: Licence agreement concerning inclusion of doctoral thesis in the Institutional Repository of the University of Leiden Downloaded from: https://hdl.handle.net/1887/18531

Note: To cite this publication please use the final published version (if

applicable).

(2)

Propositions

1. There is always a trade-off between enhancing details and reducing noise in low contrast X-ray angiographic images for visualization purposes. The ultimate goal would be to enhance details to the desired quality, while keeping the noise at an acceptable level. [This Thesis]

2. Increasing the acquisition angle difference of the two angiographic views does not increase accuracy and precision of three-dimensional quantitative coronary angiography for circular lesions or bifurcation dimensions. [This Thesis]

3. Due to the variable anatomy of each individual patient combined with the variable orientation of the heart in the thorax, the operator-selected working view can be quite different from the true optimal view. [This Thesis]

4. The anatomy-defined bifurcation optimal viewing angle cannot be obtained in roughly half of the patient population due to the mechanical constraints of the current X-ray systems. [This Thesis]

5. The discrepancy in lumen size as assessed by three-dimensional quantitative coronary angiography and by intravascular ultrasound or optical coherence tomography is more pronounced in larger and tortuous vessels. [This Thesis]

6. Visual interpretation is extremely important in medical diagnosis; however, for evidence-based medicine, quantitative image analysis takes “variability” out of the interpretation.

7. Without contributing to improved patient care, a good medical image processing method is a goal half achieved.

8. A good workflow is as important as accuracy for quantitative analytical methods that aim to be used in daily clinical practice.

9. Passion is beneath beauty, the key to beauty.

10. Innovation through observation, plus an unbounded mind.

11. The only way to be truly satisfied is to do what you believe is great work.

And the only way to do great work is to love what you do. [Steve Jobs]

Referenties

GERELATEERDE DOCUMENTEN

9  In-vivo comparison of arterial lumen dimensions assessed by co- registered three-dimensional (3D) quantitative coronary angiography, intravascular ultrasound and optical

To develop fast and reproducible approaches for the 3D X-ray angiographic reconstruction of coronary arteries including the bifurcation, and for the co-registration of X-ray

Because of the low-pass characteristics of X-ray systems, the sharpness of the visualized coronary arteries is limited (images are blurred), which become less appreciated

Given the aforementioned facts, we ignore insignificant sources of distortions and assume that the uneven gantry sag between the frontal X- ray system and the lateral X-ray

For each included matched pair, the diameter stenosis and lesion length were assessed on 3 different types of straight lesions in the brass phantom, i.e., a circular concentric

Although it was not the primary goal of this study, the finding of the correlation between the accumulated curvature and the difference in 3D QCA and IVUS segment lengths

Aims: To propose and validate a novel approach to determine the optimal angiographic viewing angles for a selected coronary (target) segment from X-ray coronary angiography,

minimal overlap and slightly more foreshortening as compared with ABOVA. However, a very important practical issue is that it can be reached by the X-ray systems. The 3D