University of Groningen

Anterior segment optical coherence tomography angiography

Ang Han Nian, Marcus

IMPORTANT NOTE: You are advised to consult the publisher's version (publisher's PDF) if you wish to cite from

it. Please check the document version below.

Document Version

Publisher's PDF, also known as Version of record

Publication date:

2019

Link to publication in University of Groningen/UMCG research database

Citation for published version (APA):

Ang Han Nian, M. (2019). Anterior segment optical coherence tomography angiography: Development and

application of OCT angiography for corneal vascularisation. University of Groningen.

Copyright

Other than for strictly personal use, it is not permitted to download or to forward/distribute the text or part of it without the consent of the author(s) and/or copyright holder(s), unless the work is under an open content license (like Creative Commons).

Take-down policy

If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim.

Downloaded from the University of Groningen/UMCG research database (Pure): http://www.rug.nl/research/portal. For technical reasons the number of authors shown on this cover page is limited to 10 maximum.

(2)

Chapter 4

Optical Coherence Tomography Angiography and Indocyanine Green

Angiography for Corneal Vascularisation

(3)

Optical coherence tomography angiography and

indocyanine green angiography for corneal

vascularisation

Marcus Ang,

1,2

Yijun Cai,

2

Becky MacPhee,

2

Dawn A Sim,

2,3

Pearse A Keane,

2,3

Chelvin C A Sng,

1,2,4

Catherine A Egan,

2,3

Adnan Tufail,

2,3

Daniel F Larkin,

2,3

Mark R Wilkins

2,3

1Singapore National Eye

Center, Singapore Eye Research Institute, Singapore, Singapore

2Moorfields Eye Hospital NHS

Foundation Trust, London, UK

3Institute of Ophthalmology,

University College London, London, UK

4Department of

Ophthalmology, National University Hospital, Singapore, Singapore

Correspondence to Dr Marcus Ang, Singapore National Eye Centre, Singapore Eye Research Institute, 11 Third Hospital Avenue, Singapore 168751, Singapore; marcus.ang@snec.com.sg Received 18 August 2015 Revised 1 December 2015 Accepted 10 January 2016

To cite: Ang M, Cai Y, MacPhee B, et al. Br J Ophthalmol Published Online First: [please include Day Month Year] doi:10.1136/bjophthalmol-2015-307706

ABSTRACT

Background/Aim To describe an optical coherence tomography angiography (OCTA) system adapted for anterior segment imaging, compared with indocyanine green angiography (ICGA) in eyes with corneal vascularisation.

Methods Retrospective study of subjects with corneal vascularisation secondary to microbial keratitis who had OCTA scans performed using a commercially available split-spectrum amplitude-decorrelation algorithm angiography system (AngioVue; Optovue Inc., Fremont, California, USA) and ICGA images (Spectralis; Heidelberg Engineering, Heidelberg, Germany). The agreement between OCTA and ICGA techniques in terms of area of vascularisation measured, using Bland–Altman 95% limits of agreement (LOA).

Results We compared the area of corneal

vascularisation in 64 scan images (eight eyes, four scans for each angiography technique). In our series, the overall mean area of vascularisation from the ICGA scans was 0.49±0.34 mm2and OCTA scans was 0.51

±0.36 mm2. We obtained substantial repeatability in

terms of image quality score (κ=0.80) for all OCTA scans. The agreement between OCTA and ICGA scans was good, although ICGA measured a smaller area compared with the OCTA with a mean difference of −0.03 mm2(95% CI−0.07 to 0.01). The LOA ranged

from a lower limit of−0.27 (95% CI −0.34 to −0.19) to an upper limit of 0.20 (95% CI 0.13 to 0.28, p=0.127).

Conclusions We found that rapid, non-contact OCTA adapted for the cornea was comparable with ICGA for measurement of the area of corneal vascularisation in this pilot clinical study. Further prospective studies are required to confirm if this relatively new imaging technique may be further developed to replace invasive angiography techniques for the anterior segment.

INTRODUCTION

Angiography for the anterior segment has a variety of clinical applications, ranging from the evaluation of corneal and scleral inflammatory disorders1 2to

guiding the treatment of corneal vascularisation.3

Fluorescein angiography (FA) and indocyanine green angiography (ICGA) have already been described for imaging both normal and abnormal corneoscleral vessels.1–5 Although fluorescein

rapidly fills and delineates vessels during the first transit of dye, subsequent leakage through the normal fenestrated capillaries in the conjunctiva

prevents visualisation of any deeper vessels.4 On

the other hand, indocyanine green (ICG) is primar-ily bound by proteins and remains in the vessels for a longer time and leaks only in the presence of damage or inflammation.5 Despite the potential

advantages and clinical indications of both techni-ques, the clinical use of these angiography methods has been limited due to their invasive nature and infrequent but potentially severe, adverse events.6

Recently, optical coherence tomography angiog-raphy (OCTA) has been used to successfully delin-eate vessels within the retina, choroid and the vasculature associated with the optic nerve.7–11

These non-contact imaging systems detect phase variations or changes in reflectivity to detect vascu-lar flow, with the added benefit of concurrently obtaining optical coherence tomography (OCT) scans of the surrounding tissue.12 This allows for

rapid, non-invasive evaluation of ocular pathologies with their associated vessels, exemplified by OCTA assessment of choroidal neovascularisation with the evaluation of surrounding choroidal and retinal layers simultaneously.13In particular, split-spectrum

amplitude-decorrelation angiography (SSADA) improves the signal-to-noise ratio of flow detection and has been shown to be useful for visualising the microvasculature networks within the eye.12

Since current OCTA systems are optimised for the retina and optic disc, we had previously described a technique adapted to perform scans in the anterior segment for normal corneal and limbal vessels.14As objective evaluation of corneal

vascu-larisation is becoming increasingly important for its diagnosis and monitoring, new non-invasive imaging modalities are a welcome addition to our clinical practice.15 However, while the OCTA

system rapidly produced angiograms of the corneal and limbal vasculature with substantial consist-ency,14 its role in evaluating abnormal corneal

vessels in comparison with current angiography techniques has not been established. Thus, we con-ducted this pilot clinical study to evaluate the same OCTA technique adapted for cornea and compared it with anterior segment ICGA—reported to be more useful than FA for delineating pathological corneal vessels.16

MATERIALS AND METHODS

In this observational study, we conducted a retro-spective review of subjects with corneal vascularisa-tion at Moorfields Eye Hospital from 1 December Ang M, et al. Br J Ophthalmol 2016;0:1–7. doi:10.1136/bjophthalmol-2015-307706 1

Clinical science

BJO Online First, published on January 28, 2016 as 10.1136/bjophthalmol-2015-307706

Copyright Article author (or their employer) 2016. Produced by BMJ Publishing Group Ltd under licence.

group.bmj.com on February 4, 2016 - Published by

http://bjo.bmj.com/ Downloaded from

(4)

2014 to 30 May 2015. We included patients with corneal vascu-larisation secondary to microbial keratitis; extending more than 3 mm from the limbus into the cornea with significant corneal lipid keratopathy and scarring; which did not respond to topical steroid therapy for at least 12 weeks and were planned to undergo fine-needle diathermy. Eyes with active inflammation, infection or ulceration were excluded. After the specific areas of corneal vascularisation for study were identified, photographs were taken using digital slit-lamp cameras (Topcon ATE-600, with D1X digital camera, Nikon Corp) with a standard diffuse illumination (×10 magnification, flash power 4, 45° angle). Our study followed the principles of the Declaration of Helsinki .

Angiography technique

A standard technique was used for anterior segment ICGA using the Heidelberg scanning laser ophthalmoscope (Heidelberg Engineering, Heidelberg, Germany) in ‘Hi-Res Mode’, using a 20° field of view with the focus setting on ‘+32.00 D’. Infrared images were captured before slow injection of 5 mL of ICG dye (5 mg/mL at 1 mL/s) and the video was recorded for 30 s, fol-lowed by single-frame ICGA images of the whole cornea every 10 s for 3 min. Four of the ICGA latest frames were chosen once the image was stable with no increase or visible leakage. All eyes then had four 6×6 mm scans of the identified area of

corneal vascularisation on a separate day at least 1 week apart, using the AngioVue OCTA system (Optovue Inc., Fremont, California, USA) with the long corneal adaptor module (CAM-L) using a previously described technique.14 In the

OCTA system used for this study, the autofocus function had to be deactivated, and the lens moved very close (2–3 cm) to the corneal surface before fine-tuning and manual adjustments to the focal lengths were made to achieve adequate focus on the area of interest in the cornea. Scans had transverse resolution of 15 μm and axial resolution of 5 μm using a light source centred on 840 nm with the beam width of 22 μm. Coronal or ‘en face’ OCTA scan images were reconstructed from 304×304 A-scans captured at 70 000 scans per second.17

Image analysis

All OCTA and ICGA images were exported from the system as a portable network graphics image file into the National Institutes of Health Image J 1.38X (National Institutes of Health, Bethesda, Maryland, USA) software for analysis using a previously described method (figure 1).14 Briefly, the ‘adjust

threshold’ function set to ‘default, HSB’ was first applied to reduce the surrounding noise and highlight the blood vessels by a single analyst masked to the diagnosis and slit-lamp photo-graphs.18 The region of interest (ROI) was encircled (radius

Figure 1 Examples of image analysis comparing indocyanine green angiography (ICGA) frames and optical coherence tomography angiography (OCTA) scans. (A and B) The area of corneal vascularisation in the cornea was first identified from ICGA frame and OCTA scan. (C and D) Examples of one ICGA frame and OCTA scan image, which were exported from the system as a portable network graphics image file into the ImageJ software (V.1.49p, Wayne Rasband; National Institutes of Health, Bethesda, Maryland, USA). The ‘adjust threshold’ function in ImageJ was used to highlight the blood vessels and reduce the surrounding noise. The area within the region of interest of 2.5 mm radius was calculated from the binary image and the total area of vessels was calculated, for example, in these images the ICGA frame (0.64 mm2) was compared with the OCTA scan image

(0.69 mm2).

2 Ang M, et al. Br J Ophthalmol 2016;0:1–7. doi:10.1136/bjophthalmol-2015-307706

Clinical science

group.bmj.com on February 4, 2016 - Published by

http://bjo.bmj.com/ Downloaded from

(5)

2.50 mm, threshold 0.05) using the in-built ‘Angioflow’ soft-ware, which identified the appropriate corneal vessels and the limbus estimated from the marginal corneal vascular arcade in the corresponding area of the corneal vascularisation of all OCTA images.19Next, we used a selective filter to produce the

vessels as a binary image, with pixel resolution (mm/pixel) for ICGA images defined as the ratio between the diameter of the cornea (mm) and the image diameter of the cornea in pixels; and diameter:pixel resolution was determined from the 2.50 mm circle on OCTA scans.20 The resulting binary image

was used to calculate the area of corneal vessels within the ROI in mm2(‘1’ for black pixels over the vessels, ‘0’ for the white

background). We assessed the image quality score using a recog-nised system, that is, 0–4 (0, no vessel discernible; 1, poor vessel delineation; 2, good vessel delineation; 3, very good vessel delineation; 4, excellent vessel delineation), performed by two independent masked assessors.16We also documented the signal

strength of all OCTA scans as determined by the AngioVue system (Optovue Inc.).

Statistical analysis

We analysed all scan images obtained for repeatability, image quality and vessel area measurements between angiography tech-niques. The area of corneal vascularisation measured from ICGA and OCTA images was assessed using the method described by Bland and Altman, that is, 95% limits of agreement (LOA) and mean difference ±1.96 SD with its 95% CIs calcu-lated with the derived p value.21We calculated the kappa

coeffi-cient (κ) value for the repeatability of scans using the image quality score, where κ≤0.2 was considered slight, 0.21–0.40 weak, 0.41–0.6 moderate, 0.61–0.8 substantial and 0.81–1.0 was considered ‘almost perfect’ in agreement.22 A p value

<0.05 was considered statistically significant for comparisons between two quadrants. Statistical Package for the Social Sciences V.17.0 (SPSS Inc., Chicago, Illinois, USA) was used to analyse the data.

RESULTS

In this study, we analysed a total of 64 scans, that is, four OCTA images and four ICGA frames in each of the eight eyes that were included. The mean age of our subjects was 45 ±0.5 years and five (62.5%) were men. All eyes had significant corneal vascularisation extending more than 3 mm from the limbus into the paracentral cornea with a mean duration of 17 ±3 months and median visual acuity of 20/50 (20/30–20/80). All scans were performed with substantial repeatability in terms of image quality score (κ=0.80) with a mean image ICGA score of 2.7±0.5, and mean image OCTA score of 2.5±0.6 with mean signal strength of 38±14. The overall mean area of vascularisation from the OCTA scans was 0.51±0.36 mm2and

ICGA scans was 0.49±0.34 mm2. There was no statistically

sig-nificant difference in the mean measurements from eight eyes comparing both techniques (p=0.674; mean difference of −0.01 mm2, 95% CI −0.11 to 0.08). Overall, the analysis of

64 scans revealed a consistent mean difference of −0.03 mm2

(95% CI −0.07 to 0.01) where the ICGA measured a smaller area compared with the OCTA. We found a good agreement between the two techniques (p=0.127) as the LOA ranged from a lower limit of −0.27 mm2 (95% CI −0.34 to

−0.19 mm2) to an upper limit of 0.20 mm2(95% CI 0.13 to

0.28 mm2) (figure 2).

In order to highlight the differences in images obtained from each angiography technique, we present an example of an eye with herpetic keratitis and secondary lipid keratopathy, whose

OCTA and ICGA images are also shown in figure 1. As his corneal scarring and vascularisation did not respond to a 6-month course of topical steroid treatment, angiography was performed to delineate the corneal vessels in preparation for fine-needle diathermy. The early ICGA scans (figure 3A,B) demonstrated the afferent vessels within the first 15–30 s, while the later scans demonstrated the remaining efferent vessels at 45 s (figure 3C). There was no leakage detected in the late frames such as that seen at 3 min (figure 3D).Figure 4 demon-strates the use of OCTA images in the same eye, which allows the depth of the corneal vessels to be assessed.

Our study also allowed us to examine some limitations of this relatively new OCTA system, originally designed for the retina. An important observation to highlight is that while the OCTA was able to image most of the vessels in the areas of corneal vascularisation, on closer examination, vessels with seg-ments or branches with a predominant flow in the axial direc-tion may not be delineated as well. We illustrate this characteristic using the eye with herpetic keratitis with a rela-tively large area of lipid deposition and corneal scarring (figure 5A,B). A single late-frame ICGA image (figure 5C) provides a 2D transverse view of the abnormal corneal vessels with no information on the depth or axial direction of each branch. A ‘close-up’ 3×3 mm OCTA scan of the same region (figure 5D) still suffers from motion artefacts, limited image resolution and a ‘gap’ in the main branch of afferent vessel as the ‘missing’ segment actually loops in the axial direction (anterior to poster-ior towards the viewer) before continuing down in a transverse direction.

DISCUSSION

Our preliminary study using a relatively new OCTA imaging technique suggests that the measurements of corneal vascular-isation are generally of good agreement with ICGA images, with at least 95% of measurements and differences within the 95% prediction intervals and 95% LOA, respectively. However, improvements to image processing and optimisation of the OCTA for the anterior segment may be required in the future developments of OCTA systems. The observed underestimation Figure 2 Bland–Altman plot comparing indocyanine green

angiography (ICGA) frames and optical coherence tomography angiography (OCTA) scans. The measured area of corneal vessels within the region of interest from the ICGA and OCTA scans (32 images each) performed in each eye was compared using the Bland–Altman plot. We found a good agreement between OCTA and ICGA scans, although ICGA underestimated the area with a mean difference of−0.03 mm2

(95% CI−0.07 to 0.01).

Ang M, et al. Br J Ophthalmol 2016;0:1–7. doi:10.1136/bjophthalmol-2015-307706 3

Clinical science

group.bmj.com on February 4, 2016 - Published by

http://bjo.bmj.com/ Downloaded from

(6)

of corneal vessel area measured by the ICGA compared with OCTA in our study is probably of minimal clinical relevance— but should be taken into consideration and reconfirmed in

future studies planning to use these imaging modalities in evaluating new treatments for corneal vascularisation. The small (0.03 mm2), albeit consistent, difference we observed,

Figure 3 Examples of indocyanine green angiography (ICGA) frames from Case 1 (figure 1). (A) 15 s. (B) 30 s. (C) 1 min. (D) 3 min. Early ICGA

frames identified the afferent vessels within the first 15 s (A), extending into the cornea scar (B). The efferent vessels and capillary loops within the corneal scar become evident at 30–45 s (C). Of note, no leakage was detected in the late frames up to 3 min (D).

Figure 4 Serial optical coherence tomography angiography (OCTA) scans from Case 1 (figure 1). Cross-sectional OCTA scans of the corneal scar and vessels at varying corneal and scleral depths. ‘Depth-sectioning’ allows for simultaneous assessment of the structure, depth of corneal stromal invasion and the abnormal blood vessels in this lesion. Estimated depth—(A) 0–50 μm; (B) 100–150 μm; (C) 200–250 μm; (D) 300–350 μm; (E) 400–450 μm; (F) 500–550 μm. Scans on the left of each frame (A–F) demonstrate the coronal or ‘en face’ view at varying depths, while the scans on the right of each frame show the traditional cross-sectional view of the cornea, which provides the depth of scan obtained within the cornea scar. Note that the segmentation lines (red and green) are not perfectly parallel with the corneal curvature, leading to several vessels being visible in overlapping cross-sectional segmental scans.

4 Ang M, et al. Br J Ophthalmol 2016;0:1–7. doi:10.1136/bjophthalmol-2015-307706

Clinical science

group.bmj.com on February 4, 2016 - Published by

http://bjo.bmj.com/ Downloaded from

(7)

that is, ICGA underestimating the area of vascularisation, may be due to fundamental differences in the imaging techniques or minor disparities that have arisen from the image analysis— such as non-parallel segmentation or projection artefacts, which may cause a superficial vessel to appear thicker than they should be.12 Our study also suggests that while the SSADA

used in this study reduced noise from the images without com-promising the flow signal, in the cornea this OCTA system may suffer from image artefacts from light scatter due to corneal scars, which may overestimate the area of vascularisation or lose detection of segments in the vessel with minimal flow or axial flow—as the algorithm used is optimised for the retina and choroid, where vessels exhibit mainly transverse flow.12

However, the aim of this early clinical study was not to deter-mine whether either imaging technique was superior or why these observed differences exist—but instead, to demonstrate the usefulness of this fairly unproven OCTA retinal imaging system adapted for the cornea, and how the measurements compare with the more widely accepted ICGA technique in our current clinical practice.

This study also reinforced several advantages, as well as the current limitations of OCTA adapted for the anterior segment

and cornea, compared with an established angiography tech-nique such as ICGA. The first and most obvious advantage of the OCTA is its ability to perform non-contact, cross-sectional and coronal sections of areas of interest in the cornea and asso-ciated vessels within seconds.14However, the field of view and

image resolution are currently limited, with several image arte-facts, which may be due to the presence of scarring, and the coronal reconstruction of OCT scans, as demonstrated in

figures 4and 5. While clinicians may make up for this by per-forming multiple scans and ensuring an acceptable signal strength (as determined by the in-built ‘Angioflow’ system in this study), improvements to the software and optimisation for the anterior segment may further improve the field of view and image resolution—before which a clinician may choose to perform ICGA in eyes with concomitant dense corneal scarring.

Second, the OCTA is able to provide high-resolution scans of the cornea with accompanying information on the depth of abnormal vasculature—useful for evaluating corneal scars and any associated vessels while planning for procedures such as lamellar keratoplasty and fine-needle diathermy,3 or evaluation

of peripheral corneal infiltrates or melts with the adjacent Figure 5 Figures to demonstrate a closer examination of slit-lamp images, indocyanine green angiography (ICGA) frames and optical coherence tomography angiography (OCTA) scans in another example of corneal vascularisation. In this example of an eye with lipid keratopathy and corneal vessels (A), a higher-magnification of the slit-lamp image reveals the abnormal vasculature (B). The late ICGA frame taken at 3 min confirms the branching patterns of the pathological vessels in a 2D view without stereoscopic appreciation of the depth or direction of vessels (C). Multiple 3×3 mm OCTA scans of good signal strength were taken of the same area and consistently showed a gap in the main vessel (D) as the segment loops from a transverse to axial direction, before extending further down. As this segment of the afferent vessel was superficial, the discontinuity of the vessel shown on OCTA is unlikely to be secondary to minimal flow or blocked signal from the overlying lipid deposition, but may be due to the OCTA being less sensitive to axial flow.

Ang M, et al. Br J Ophthalmol 2016;0:1–7. doi:10.1136/bjophthalmol-2015-307706 5

Clinical science

group.bmj.com on February 4, 2016 - Published by

http://bjo.bmj.com/ Downloaded from

(8)

inflamed sclera and limbal vessels.14 However, unlike the FA

or ICGA, the OCTA is unable to demonstrate the sequence of flow, that is, the afferent versus efferent vessels, and is unable to demonstrate any vessel leakage or presence of lymphatics. Leakage of fluorescein occurs rapidly during anterior segment FA through normal vessels, which may actually obscure the true extent of corneal vascularisation.16 While the presence of

fluorescein or ICG leakage influences clinical management for retinal or choroidal pathology, in the anterior segment leakage that suggests inflammation may add limited clinical informa-tion in the presence of already inflamed sclera.4On the other

hand, the absence of limbal vessels may be a more useful clin-ical sign of severity in corneal melting disorders and periph-eral ulcerative keratitis—a sign that is often obscured by the leakage or extravasation of dye.2 The progression of corneal

melting and the need for systemic immunosuppression in such severe inflammatory conditions are usually preceded by vasculitis and vaso-occlusion of the limbal vessels, while recanalisation and new capillary formation may indicate response to treatment.2 Thus, OCTA has the potential to play

an important role in detecting progression, prognostication and management of these corneoscleral destructive diseases, which requires further studies in these specific conditions to confirm.

Lastly, our study revealed that performing the OCTA, by trained technicians as well as clinicians, had a relatively easy learning curve. As this is a non-invasive procedure that requires minutes to perform multiple scans (3–4 s per scan) with substan-tial consistency, patients are not exposed to the potensubstan-tial adverse reactions associated with the administration of intraven-ous dyes. Significant time and preparation are required before each ICGA or FA imaging session, while some patients may not be suitable for this procedure at all, due to various contraindica-tions.23 24However, it must be emphasised that current OCTA

systems are not optimised for the anterior segment, and adjust-ments had to be made to the scanning technique while using the lens adapter.

The potential clinical applications of OCTA for the anterior segment, apart from those already mentioned, could also extend to addition to intra-operative OCT assessment,25

assess-ment of graft vascularisation with prognostication for graft rejection,26 evaluation of new antiangiogenic treatments for

corneal vascularisation,27 studying limbal vasculature associated

with limbal stem cell deficiency28 or even evaluation of bleb

vascularity and morphology after glaucoma surgery.29 We

rec-ognise the limitations of this preliminary study, which describes the adaptation of the retinal OCTA compared with ICGA for the evaluation of corneal vascularisation. Ideally, a large pro-spective study of various pathological corneal vessels with com-parisons with slit-lamp photography and angiography should be performed to evaluate this OCTA technology specifically for the anterior segment. It would also be useful to perform serial OCTA scans on the same eyes to monitor progression or treat-ment outcomes. Nonetheless, we present observations from an early clinical study that objectively discusses the advantages and limitations of this novel OCTA system, which may develop further to fulfil its potential as a non-invasive alternative to current angiography techniques. In its current iteration, our findings suggest the OCTA system used in this study may be useful for monitoring areas of corneal vascularisation, but less effective for identifying afferent vessels to guide targeted therapy. Further research using OCTA systems optimised for the anterior segment in various other corneal vascular

pathologies would be required to support the observations from this study.

Twitter Follow Mark Wilkins at @wilkoman

Funding PAK, DAS, CAE, AT, DFL and MRW have received a proportion of their funding from the Department of Health’s National Institute for Health Research (NIHR) Biomedical Research Centre for Ophthalmology at Moorfields Eye Hospital and UCL Institute of Ophthalmology.

Competing interests AT is a member of the Optovue Inc. users group that received non-financial, technical support.

Patient consent Obtained.

Ethics approval Moorfields Eye Hospital NHS Trust and local institutional review board.

Provenance and peer review Not commissioned; externally peer reviewed. Data sharing statement Any additional unpublished data may be obtained from the corresponding author.

REFERENCES

1 Watson PG, Bovey E. Anterior segment fluorescein angiography in the diagnosis of scleral inflammation.Ophthalmology1985;92:1–11.

2 Watson PG. Anterior segment fluorescein angiography in the surgery of immunologically induced corneal and scleral destructive disorders.Ophthalmology 1987;94:1452–64.

3 Spiteri N, Romano V, Zheng Y, et al. Corneal angiography for guiding and evaluating fine-needle diathermy treatment of corneal neovascularization. Ophthalmology2015;122:1079–84.

4 Guex-Crosier Y, Durig J. Anterior segment indocyanine green angiography in anterior scleritis and episcleritis.Ophthalmology2003;110:1756–63. 5 Nieuwenhuizen J, Watson PG, Emmanouilidis-van der Spek K, et al. The value of

combining anterior segment fluorescein angiography with indocyanine green angiography in scleral inflammation.Ophthalmology2003;110:1653–66. 6 Kirwan RP, Zheng Y, Tey A, et al. Quantifying changes in corneal neovascularization

using fluorescein and indocyanine green angiography.Am J Ophthalmol 2012;154:850–8.e2.

7 Ishibazawa A, Nagaoka T, Takahashi A, et al. Optical coherence tomography angiography in diabetic retinopathy: a prospective pilot study.Am J Ophthalmol 2015;160:35–44.e1.

8 Spaide RF, Klancnik JM Jr, Cooney MJ. Retinal vascular layers imaged by fluorescein angiography and optical coherence tomography angiography.JAMA Ophthalmol 2015;133:45–50.

9 Schwartz DM, Fingler J, Kim DY, et al. Phase-variance optical coherence tomography: a technique for noninvasive angiography.Ophthalmology2014;121:180–7. 10 Jia Y, Wei E, Wang X, et al. Optical coherence tomography angiography of optic

disc perfusion in glaucoma.Ophthalmology2014;121:1322–32. 11 Jia Y, Bailey ST, Wilson DJ, et al. Quantitative optical coherence tomography

angiography of choroidal neovascularization in age-related macular degeneration. Ophthalmology2014;121:1435–44.

12 Jia Y, Tan O, Tokayer J, et al. Split-spectrum amplitude-decorrelation angiography with optical coherence tomography.Opt Express2012;20:4710–25.

13 de Carlo TE, Bonini Filho MA, Chin AT, et al. Spectral-domain optical coherence tomography angiography of choroidal neovascularization.Ophthalmology 2015;122:1228–38.

14 Ang M, Sim DA, Keane PA, et al. Optical coherence tomography angiography for anterior segment vasculature imaging.Ophthalmology2015;122:1740–7. 15 Cursiefen C, Colin J, Dana R, et al. Consensus statement on indications for

anti-angiogenic therapy in the management of corneal diseases associated with neovascularisation: outcome of an expert roundtable.Br J Ophthalmol2012;96:3–9. 16 Anijeet DR, Zheng Y, Tey A, et al. Imaging and evaluation of corneal vascularization

using fluorescein and indocyanine green angiography.Invest Ophthalmol Vis Sci 2012;53:650–8.

17 Spaide RF, Klancnik JM Jr, Cooney MJ. Retinal vascular layers in macular telangiectasia type 2 imaged by optical coherence tomographic angiography.JAMA Ophthalmol2015;133:66–73.

18 Sharma A, Bettis DI, Cowden JW, et al. Localization of angiotensin converting enzyme in rabbit cornea and its role in controlling corneal angiogenesis in vivo. Mol Vis 2010;16:720–8.

19 Ang M, Cai Y, Shahipasand S, et al. En face optical coherence tomography angiography for corneal neovascularisation.Br J Ophthalmol2015. Published Online First 26 Aug 2015. doi:10.1136/bjophthalmol-2015-307338

20 Zheng Y, Kaye AE, Boker A, et al. Marginal corneal vascular arcades.Invest Ophthalmol Vis Sci2013;54:7470–7.

21 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement.Lancet1986;1:307–10.

6 Ang M, et al. Br J Ophthalmol 2016;0:1–7. doi:10.1136/bjophthalmol-2015-307706

Clinical science

group.bmj.com on February 4, 2016 - Published by

http://bjo.bmj.com/ Downloaded from

(9)

22 Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics1977;33:159–74.

23 Kwiterovich KA, Maguire MG, Murphy RP, et al. Frequency of adverse systemic reactions after fluorescein angiography. Results of a prospective study. Ophthalmology1991;98:1139–42.

24 Stanga PE, Lim JI, Hamilton P. Indocyanine green angiography in chorioretinal diseases: indications and interpretation: an evidence-based update.Ophthalmology 2003;110:15–21; quiz 2–3.

25 De Benito-Llopis L, Mehta JS, Angunawela RI, et al. Intraoperative anterior segment optical coherence tomography: a novel assessment tool during deep anterior lamellar keratoplasty.Am J Ophthalmol2014;157:334–41.

26 Bachmann B, Taylor RS, Cursiefen C. Corneal neovascularization as a risk factor for graft failure and rejection after keratoplasty: an evidence-based meta-analysis. Ophthalmology2010;117:1300–5.e7.

27 Chang JH, Gabison EE, Kato T, et al. Corneal neovascularization.Curr Opin Ophthalmol2001;12:242–9.

28 Kim YJ, Yoo SH, Chung JK. Reconstruction of the limbal vasculature after limbal-conjunctival autograft transplantation in pterygium surgery: an angiography study.Invest Ophthalmol Vis Sci2014;55:7925–33. 29 Sng CC, Singh M, Chew PT, et al. Quantitative assessment of changes in

trabeculectomy blebs after laser suture lysis using anterior segment coherence tomography.J Glaucoma2012;21:313–17.

Ang M, et al. Br J Ophthalmol 2016;0:1–7. doi:10.1136/bjophthalmol-2015-307706 7

Clinical science

group.bmj.com on February 4, 2016 - Published by

http://bjo.bmj.com/ Downloaded from

Referenties

GERELATEERDE DOCUMENTEN

The main goals of my PhD thesis are (1) to establish that OCTA systems designed for the retina (posterior segment of the eye) can also be used to image the cornea

De hoofddoelen van mijn proefschrift zijn (1) om te laten zien dat OCTA-systemen die zijn ontworpen voor het netvlies (achtersegment van het oog) ook gebruikt kunnen worden voor het

He is also involved in the Leadership Development Program at the Asia Pacific Academy of Ophthalmology and American Academy of Ophthalmology, and is on several

Long-term Visual Outcomes comparing Descemet Stripping Automated Endothelial Keratoplasty and Penetrating Keratoplasty.. Sharma S, Ang M, Najjar RP, Sng C, Cheung CY, Rukmini

If you believe that this document breaches copyright please contact us providing details, and we will remove access to the work immediately and investigate your claim. Downloaded

3D angiographic reconstruction and quantitative analysis were performed by an experienced analyst using a novel and validated 3D QCA software package (prototype version, Medis

License: Licence agreement concerning inclusion of doctoral thesis in the Institutional Repository of the University of Leiden Downloaded.

9  In-vivo comparison of arterial lumen dimensions assessed by co- registered three-dimensional (3D) quantitative coronary angiography, intravascular ultrasound and optical