• No results found

A prefilled, ready-to-use, electrophoresis-based lab-on-a-chip device for monitoring ions in blood and urin

N/A
N/A
Protected

Academic year: 2021

Share "A prefilled, ready-to-use, electrophoresis-based lab-on-a-chip device for monitoring ions in blood and urin"

Copied!
3
0
0

Bezig met laden.... (Bekijk nu de volledige tekst)

Hele tekst

(1)

A PREFILLED, READY-TO-USE, ELECTROPHORESIS-BASED

LAB-ON-A-CHIP DEVICE FOR MONITORING IONS IN BLOOD AND URINE

S.S. Staal1*, J. Floris1, S.O. Lenk1, E. Staijen1, M. Avilla Muñoz2, D. Kohlheyer3, J.C.T. Eijkel3, and

A. van den Berg3

1Medimate BV, THE NETHERLANDS, 2University of Castilla-La Mancha, SPAIN,

3MESA+ Institute for nanotechnology,University of Twente, THE NETHERLANDS

ABSTRACT

We present the Medimate Multireader®, the first point-of-care Lab-on-a-Chip device based on Capillary

Electropho-resis. We show that it is possible to analyze lithium in whole blood and sodium in urine, using a commercially viable Multireader that is suitable for self testing. In addition, we demonstrate the ability to detect creatinine using the same platform.

KEYWORDS: Lab-on-a-Chip, Point-of-Care, Capillary Electrophoresis, Conductivity Detection, In Vitro Diagnostics, lithium, sodium, creatinine

INTRODUCTION

At present, most diagnostics in the clinical laboratory are carried out on large clinical devices, only to be used by trained personnel. It was expected that microfluidics would change the laboratory setting, because of the miniaturization possi-bilities of the lab-on-a-chips. Until now the iStat has shown to be viable as a commercial product, but operation is still limited to professionals. Research at the University of Twente, including commercialization, has now resulted in the first Microchip Capillary Electrophoresis (MCE) platform suitable for analysis of electrolytes. Most importantly, it can be used by lay persons in their home environment.

Research at the Mesa+ Institute for Nanotechnology, performed from 2000 to 2005, proved the possibility to measure

li-thium in a whole blood sample.1,2,3 Between 2005 and 2010, further development by Medimate resulted in a

commercial-ly viable in vitro diagnostic device for lithium measurements that is suitable for self testing, the Medimate Multireader®. The system consists of a disposable µCE chip and the Multireader, see Figure 1.

Figure 1. Elektrolyte measurement: a) shipped blister including the prefilled µCE chip, assembled in a plastic disposa-ble for ease of handling and sample application b) application of a single blood droplet from finger stick c) disposadisposa-ble inserted into the Multireader, including high voltage power supply and conductivity detection unit. The measurement is completed typically within 1 to5 minutes.

During the commercialization phase, many challenges had to be overcome. For instance, lithium, the first analyte that was targeted, is one of the most difficult electrolytes to be analyzed. Lithium is used as a drug for patients with Manic Depressive Illness, also known as Bipolar Disorder. It is therapeutic in the range of 0.4 – 1.2 mmol/l. The technological challenge with lithium is the low concentration range, compared to other electrolytes, like potassium, calcium, magne-sium and particularly sodium. The detection limit of lithium is determined to be 0.2 mmol/l with an accuracy of 0.1 mmol/l or 10 %. This detection limit is 700 fold lower than the normal sodium concentration. To be able to measure the lithium accurately, a very high resolution and a very high signal to noise ratio had to be realized. This was achieved by many chemical, electrical and electrochemical innovations. Amongst others, it was chosen to use a novel single entrance solution, combined with a furthermore closed embodiment to perform the MCE process.

In addition, the shelf life was addressed. To prevent leaking of the closed embodiment, which is designed as a commer-cially viable prefilled disposable with a sufficient shelf life, the pressure gradients that are caused by expan-sion/shrinkage of the fluid due to temperature changes had to be solved. For this specific problem Medimate has incorpo-rated a gas filled expansion chamber inside the µCE chip. Here we describe the progress of the Medimate Multireader ® as an in vitro diagnostic device to test lithium, sodium and creatinine.

978-0-9798064-3-8/µTAS 2010/$20©2010 CBMS 2107 14th International Conference on Miniaturized Systems for Chemistry and Life Sciences 3 - 7 October 2010, Groningen, The Netherlands

(2)

THEORY

To perform a measurement the sample is applied at the entrance of the channel system. The electrolytes from the sample are injected in the back ground electrolyte by moving boundary electrophoresis after applying an electrical vol-tage across the sample and channel system. In the double-T an injection plug is formed. Changing the high volvol-tage po-tential results in the injection of this plug into the separation channel by capillary zone electrophoresis. At the end of the channel the in contact conductivity detection electrodes measure the change in conductivity. The time at which the dif-ferent species pass the conductivity electrode is an indication for their identity and the change in conductivity is a meas-ure for their concentration.4,5,6,7

EXPERIMENTAL

In Figure 2 a picture and schematic representation of the µCE chip are shown. To finalize the product the chip is pre-filled with the background electrolyte and assembled in the housing prior to blistering. Combined with the Multireader different serum, whole blood and urine samples can be measured.

Figure 2 Left: photograph of the microfluidic chip; right: Schematic indication of the different functional units (the exact locations slightly differ from the photograph for the sake of clarity). 1) sample opening with applied sample

drop-let; 2) evaporation reservoir; 3) injection channel for injection of cat ions by moving boundary electrophoresis; 4) double-T injector; 5) reservoir with gas bubble for liquid expansion control; 6) conductivity detection electrodes; A, B) high-voltage injection anode and cathode; C,D) high-voltage separation anode and cathode; E) Electrodes for the

con-ductometric determination of the sample conductivity.

RESULTS AND DISCUSSION

At present the Medimate Multireader® has been validated for lithium tests in serum and whole blood and received a CE 0344 certificate for self testing. This validation is founded on the results of serum measurements and whole blood measurements, as are shown in Figure 3. The results are within the medical relevance region. Currently market tests are being performed. 0 0,2 0,4 0,6 0,8 1 1,2 1,4 0 0,2 0,4 0,6 0,8 1 1,2 1,4 y = 0,031285 + 0,96379x R= 0,95522

[Li] / mM - flame photometer

[Li] / mM - M ultir ea der 0 0,2 0,4 0,6 0,8 1 0 1 2 3 4 5 6 7 8 9 10 11

IL943 serum before IL943 serum after MR venous before MR venous after MR capillary L ith iu m co nce n tr at io n / mM Patient number

Figure 3, Left: Comparison of Multireader measurements with flame photometer IL943 measurements, both in se-rum. The measurement range is valid up to 7 mmol/l (obtained from spiked samples). Right: results of a measurement series on 10 patients. First, lithium concentrations were determined in serum by the flame photometer (IL943 serum

be-fore) and in venous blood by the Multireader (MR venous bebe-fore). Next, during approximately 100 minutes lithium was measured in capillary blood (average of 40 measurements, presented as MR capillary). Finally lithium was determined

in serum (IL943 serum after) and in venous blood (MR venous after).

Recent tests have also proven the possibility to measure sodium in urine. Figure 4 shows the results of a comparison of internal tests with the Flame Photometer. At this moment validation tests are performed at the Medisch Spectrum Twente Hospital in Enschede, the Netherlands. A specific challenge with the measurement of sodium in urine is the high dynamic range of species within the sample. To handle different stacking effects an internal standard is used, which is added to the urine sample prior to measurement.

. 1 3 E 5 6 2 A B C D 4 2108

(3)

Preliminary tests have demonstrated the possibility to detect creatinine at 1 mmol/l in serum. Future research is fo-cused on improvements on the detection limit of creatinine and the reproducibility of the measurements.

Figure 4, Left: Comparison of sodium in urine measurements using the Medimate Multireader® and the flame pho-tometerIL943. Right: Indicative creatinine measurement. The serum sample was spiked with 1 mM creatinine, which can

be observed in the electropherogram.

CONCLUSION

The Medimate Multireader® with the µCE is a capillary electrophoresis platform capable of measuring electrolytes in a commercially viable manner. Validation results show the possibility to detect lithium in serum and whole blood in the therapeutic and toxic range of the medication. Furthermore, first results are shown for the measurement of sodium in urine and the possibility of detection of creatinine using the same platform.

FUTURE DEVELOPMENTS

Medimate will extend the lithium analysis tests in the market and start rolling out the system in Europe. Furthermore the CE platform will be adapted for the detection of more species, starting with sodium and creatinine. Medimate will also continue the development of a smaller design of the Multireader, as is shown in figure 5.

Figure 5: Smaller model of the Medimate Multireader ®, expected 2011.

ACKNOWLEDGEMENTS

The STW, Agentschap NL, Stichting Achmea Gezondheidszorg and het Europees fonds voor regionale ontwikkeling from the European Unionare gratefully acknowledged for their financial support. The Medisch Spectrum Twente and Pieternel Kölling are gratefully acknowledged for their organizational support.

REFERENCES

[1] E.X. Vrouwe, R. Luttge, and A. van den Berg, Direct measurement of lithium in whole blood using

microchip capillary electrophoresis with integrated conductivity detection, Electrophoresis 2004, 25, 1660-1667 [2] E.X. Vrouwe, R. Luttge, W. Olthuis and A. van den Berg, Microchip analysis of lithium in blood using moving boundary electrophoresis and zone electrophoresis, Electrophoresis 2005, 26, 3032–3042

[3] E.X. Vrouwe, R. Luttge, I. Vermes, and A. van den Berg, Microchip Capillary Electrophoresis for Point-of-Care Analysis of Lithium, Clin. Chem. 2007, 53, 117-123

[4] J. Floris, S.S. Staal, S.O. Lenk, E. Staijen, D Kohlheyer, J.C.T. Eijkel, A. van den Berg, A prefilled, ready-to-use electrophoresis based Lab-on-a-Chip device for monitoring lithium in blood, Lab on a Chip 2010, 10, 1799-1806 [5] European patent applications EP2150815, [6] European patent application filed, [7] European patent application EP2089699

CONTACT

* Steven S. Staal, steven.staal@medimate.com

Referenties

GERELATEERDE DOCUMENTEN

Of het MVO-gegeven bruikbaar is voor een ongevallenanalyse wordt voornamelijk bepaald door de kwaliteit van dit gegeven én het aantal ongevallen (= kwantiteit)

Deze benaderingen sluiten meer aan bij de opzet voor het prognosemodel, omdat alleen op die wijze de fasering in het model (mobiliteit ~ ontmoe- tingen ~ ongevallen ~

afdeling der elektrotechniek • groep elektromechanica rapport nr. on the inforamtion electronica. The power electronic part of the system for P-regulation of the

In this thesis the theory of the deflection aberrations will first of all be con- sidered. It will be shown that the error expressions can be obtained from very

Thus, he/she always holds the other ear upwards to hear the voice of the Spirit, to discern the message of the Gospel, and in order to hear and dis- cern the voices coming from

The narrative of the two brothers indicated that they experienced divine discomfort within their relationship with the other, based on their ethical consciousness,

A convenience sample of four free-of-charge websites was evaluated on the criteria of the HON Code of Conduct for Medical and Health Websites (Lunt and Carrera 2011), which

En outre, on verra si les contraintes que Poplack propose pour l’alternance codique (la Contrainte du Morphème Libre et la Contrainte de l’Équivalence) valent aussi pour le