Quantitative
photoacoustic
integrating
sphere
(QPAIS)
platform
for
absorption
coefficient
and
Gru¨neisen
parameter
measurements:
Demonstration
with
human
blood
Yolanda
Villanueva-Palero,
Erwin
Hondebrink,
Wilma
Petersen,
Wiendelt
Steenbergen
*
BiomedicalPhotonicImagingGroup,MIRAInstituteforBiomedicalTechnologyandTechnicalMedicine,UniversityofTwente,POBox217,7500AEEnschede, TheNetherlands
1. Introduction
Quantitativephotoacousticimaging(QPAI)inbiomedicineaims
at determining target chromophore concentrations such as
endogenoushemoglobinin humanbloodorexogenous contrast
agentlevels[1].Accuratemeasurementofconcentrationscanbe
obtainedfromtheabsorptioncoefficient
m
aofabsorbersofknownmolarextinctioncoefficients.InPAimages,
m
acanbeaccuratelymeasured if the Gru¨neisenparameter
G
of thetargetchromo-phoresisknown [1]. Currentphotoacousticimaging techniques
estimatetheinitialpressuredistribution
s
owhichisaproductofthesefactorsandthefluencedistribution
F
:s
o=Gm
aF
[1].Accu-ratemeasurementofeachfactorcangiveagoodestimationof
s
owhichcanleadtocontrastonphotoacousticimageswhichhasa
quantitative interpretation. Reconstruction algorithms usually
assumeaconstant
G
foralltargetabsorberswhereinmeasurementof
s
o mainly indicates the absorbed optical energy densitym
aF
.However,differentmaterialshavedifferentG
valuessinceG
is also directly related to the material’s thermodynamicpropertiessuchasthermalexpansioncoefficient
b
,specificheatCp,andspeedofsound
v
s,asitisdefinedasG
bv
2s=Cp[2,3].Forexample,inbiologicaltissues,
G
variesfromaround0.14forblood[4]to0.80forfat[5].Severalpublicationshavereporteddifferent
techniquestomeasure
G
ofbiologicalchromophores[4–6].How-ever,theexperimentalsetupsrequireabsolutedetectionsensitivity
measurements of the optical and acoustic signals and involve
stringentalignmentbetweentheincidentlightandtargetabsorber
and acoustic detector which may not be very convenient for
measuringwithliquidsamples.Inthispaper,wepresentamethod
forsimultaneouslymeasuring
m
aandG
ofsmallvolumesofabsorbingandscatteringliquidsinjectedina softtransparenttubemounted
through two integrating spheres [7]. With integratingspheres as
platform,uniformilluminationon thetargetabsorberisachieved.
Measuring
m
aofabsorbingsamplesinatubeinsideanintegratingsphereispossibleeveninthepresenceofscattering.Thismethodof
determining
m
aiscombinedwiththetechniqueformeasuringG
oftarget absorbers in photoacoustic setup. The coupled integrating
sphere system is referred to as the quantitative photoacoustic
integratingsphere(QPAIS).Anequationformeasuring
m
ausingtheintegratingsphereisderived.Detailsoftheexperimentalsetupand
proceduresareenumerated.Absolutemagnitudesofopticalenergy
andpressurearenotnecessaryfordetermining
m
aandG
;insteadaninsitucalibrationofthesystemisdonepriortomeasurementwith
samples ofinterest. The use ofthe system is demonstratedwith
measurementsonhumanbloodsamples.Measurementsaredoneat
roomandbodytemperaturesusinganincubator.
ARTICLE INFO Articlehistory:
Received31March2016
Receivedinrevisedform24February2017 Accepted18March2017 Keywords: Photoacoustics Absorptioncoefficient Gru¨neisenparameter Integratingsphere Quantitativephotoacoustics Humanblood ABSTRACT
Quantitative photoacoustic imaging in biomedicine relies on accurate measurements of relevant materialpropertiesoftargetabsorbers.Here,wepresentamethodforsimultaneousmeasurementsof theabsorptioncoefficientandGru¨neisenparameterofsmallvolumeofliquidscatteringandabsorbing mediausingacoupled-integratingspheresystemwhichwerefertoasquantitativephotoacousticintegrating sphere(QPAIS)platform.Thederivedequationsdonotrequireabsolutemagnitudesofopticalenergyand pressurevalues,onlycalibrationofthesetupusingaqueousinkdilutionsisnecessary.Asademonstration, measurementswithbloodsamplesfromvarioushumandonorsaredoneatroomandbodytemperatures usinganincubator.Measuredabsorptioncoefficientvaluesareconsistentwithknownoxygensaturation dependence ofblood absorption at750nm, whereas measured Gru¨neisenparameter valuesindicate variabilityamongfivedifferentdonors.AnincreasingGru¨neisenparametervaluewithbothhematocritand temperatureisobserved.Theseobservationsareconsistentwiththosereportedinliterature.
ß2017UniversityofTwente.PublishedbyElsevierGmbH.ThisisanopenaccessarticleundertheCC BY-NC-NDlicense(http://creativecommons.org/licenses/by-nc-nd/4.0/).
* Correspondingauthor.
E-mailaddress:w.steenbergen@utwente.nl(W.Steenbergen).
ContentslistsavailableatScienceDirect
Photoacoustics
j ou rna l h ome p a ge : w ww . e l se v i e r. co m/ l oc a te / p a cs
http://dx.doi.org/10.1016/j.pacs.2017.03.004
2213-5979/ß2017UniversityofTwente.PublishedbyElsevierGmbH.ThisisanopenaccessarticleundertheCCBY-NC-NDlicense(http://creativecommons.org/licenses/ by-nc-nd/4.0/).
2. Methodology
2.1. Doubleintegratingspheremethodandexperimentalsetup
The methodfor measuring
G
of absorbing liquids using anintegrating sphere was already published [8] and was also
implemented and briefly described in this paper. The main
modificationswereas follows: (1) thecentralfrequency of the
transducerusedforphotoacousticdetectionwas5MHz,(2)asoft
transparentpolyethylenetubewithinnerdiameter=0.58mmand
outer diameter=0.96mm was used and (3) another similar
integrating sphere with the same physical properties was
connectedtothe one usedfor photoacoustic measurements so
thattheabsorptioncoefficientoftheabsorbingliquidinsidethe
tubecouldbesimultaneouslymeasuredusingspectrophotometry.
Fig.1showsaschematictopviewillustrationofthecoupled
integratingspheresetups.Forclarity,onlytheopticalsourcesand
detectorsareshown.Thetransducer(OlympusPanametricsNDT
V309 5MHZ/0.5in. 878182), positioned along the z-axis, the
amplifier(PanametricsNDTUltrasonicPreamp5678)and
oscillo-scope (200MHz, 2 GS/s, Tektronix TDS 2022C/24C) used for
photoacousticdetectionarenotshownontheillustration.Details
onphotoacousticmeasurementsaregivenin[8].
Asoftpolyethylenetube(Portex,SmithsMedicalInternational,
Ltd.,UK)withinnerdiameterof0.58mmwasinsertedthrough
smallholesoneachintegratingsphere(ThorlabsIS200)suchthat
the tube was positioned horizontally inside both spheres. The
verticalheightofthetube wasabout4mm aboveeach sphere
centerwhichensuredthatlightwasdirectlyincidentonthesphere
wall,avoidingdirectilluminationofthesample.
Absorptionmeasurementwasdoneusinganair-filled
integrat-ingsphere1.Ahalogenlamp(AvantesAvaLight-Hal)lightsource
wasfiber-coupledtothisintegratingspherewhileaspectrometer
(AvantesAvaSpec2048)connectedtoanotherportonthissphere
usingasimilarfiberopticwasusedtomonitorthevariationonthe
opticaloutputsignal.
Photoacousticmeasurementof
G
wasdoneusingintegratingsphere2whichwasfilledwithwaterforacousticmatchingwith
the transducer. A pulsed laser source (OpoletteTM 532I) with
wavelengthof750nm,pulselength7ns anda pulserepetition
frequencyof20Hzwasconnectedtothissecondsphereusingan
optical fiber (Newport, 0.22 NA, core diameter of 1mm).
Photodetectors (Thorlabs DET10A/M-Si detector) were used to
monitortherelativeinputandoutputlightenergy.Theuniform
illuminationonthetubewasnotaffectedbytheintroductionof
water inside the sphere cavity since the sphere wall coating
(spectralonmaterial)maintainsitshighreflectancepropertyeven
inthepresenceofwater.Moreover,thetubewasinsertedinavery
smallhole(withdiametersameastheouterdiameterofthetube)
throughtheappropriateportplugssuchthatwaterdidnotleakout
ofthesphere.
For body temperature measurements, the two-integrating
sphere setup was placed inside an incubator where ambient
temperature adjustments and measurements were possible. A
built-inblowerwarmedtheairwhileathermalsensorcontrolled
the temperature inside the incubator. It took approximately
10min to increase from room temperature 228C to body
temperatureof378C.Thereservoirswiththemediumofinterest
werealsoplacedinsidetheincubatorsothattheywereinthermal
equilibrium with the entire system during optical and
photo-acousticmeasurements.Anotherthermalprobe(National
Instru-ments NI USB-TC01) was used to monitor and record the
temperaturemeasurements.Immediatelybefore absorptionand
photoacoustic measurements, the air temperature within the
incubator,water temperatureinside theintegrating sphereand
temperatureofthesamplesweremeasured toensurethatthey
wereallinthermalequilibrium.Thetemperaturewasmonitored
throughouttheexperiments,withthethermalprobepositioned
closetothetubethatgoesintointegratingsphere2.
2.2. Preparationofhumanbloodsamples
Bloodsamplesfromhealthy donors wereobtained fromthe
Experimental Center for Technical Medicine (ECTM) of the
UniversityofTwente(UT),whichimplementsproperethicsand
approved procedure in utilizing humans and humantissues in
research.Bloodwasdirectlydrawnintovacuumsealedtubeswith
anticoagulant (either EDTA or heparin) for temporary storage.
Immediatelybeforeabsorptionandphotoacousticmeasurements,
bloodfromeachvacuum-sealedtubewaspipettedintoEppendorf
tubestoobtainthreesamplesfromthesamedonor,eachabout
1mlinvolume.Thisstepensuredthattheoxygensaturationofthe
hemoglobininthebloodwouldnotabruptlychangewhenblood
wasinjectedintothetubeintheintegratingspheresetup.
Severalfreshhumanbloodsampleswerecollectedonvarious
days. Measurements were done to investigate the measurable
valuesof
m
aandG
forbloodsamples(1)drawnfromthesamedonoronvariousdayswithnewsetupcalibration(2)drawnfrom
threedifferentdonorsonthesamedaywithsamesetupcalibration
and(3)drawnfromvariousdonorsonvariousdayswithnewsetup
calibration. Absorption and photoacoustic measurements were
donewithinonetotwohoursafterthebloodsamplewasdrawn
fromthe donor.Measurements on thesame dayweredone to
indicateusingthesamesetofcalibrationconstantsformeasuring
m
aandG
ofbloodsamplesfromvariousdonors.Ontheotherhand,measurements on various days implied investigation on the
measurablevalues using new calibrationmeasurements of the
systemsincecalibrationwasalwaysdone immediatelypriorto
measurementswiththesamplesofinterest.
Forhematocritdependenceinvestigation,wholebloodsamples
inside the vacuumsealed tube wereplaced in a centrifuge for
about10minat2000rpmuntilallRBCsettledintothebottomof
thetube.PlasmaandRBCwereseparatedandwerepipettedinto
Eppendorf tubes to obtainabout threesamples fromthe same
donor,eachabout1mlinvolume.Mixturesofredbloodcells(RBC)
andplasmainvaryingRBCconcentrations(forexample,30vol%
RBC tohave hematocrit of approximately 30%) wereprepared.
Actual hematocrit values were determined by measuring the
relativeheightoftheRBCcolumnincapillarytubes.
In order to have absorbing plasma samples at the 750nm
wavelength,smallamountofindocyaninegreendyesolutions(less
Fig.1.Top-viewschematicdiagramofthedoubleintegratingspheresetupfor measuringabsorption coefficientmaand Gru¨neisenparameterG. Thetubeis
positionedapproximately4mmabovethecenterplanewheretheopticalportsare located.
than10vol%)wereaddedtoplasmasamplespriortopipettinginto
Eppendorf tubes to obtain three 1ml samples from the same
donor,whichcorrespondstobloodssampleswithzerohematocrit.
2.3.
m
ameasurementEachabsorbingsampleofinterestwasinjectedintothetube
untilitflowedouttheotherendtoensurethatthesamesample
wasmountedinbothintegratingspheres.Simultaneousdetection
oftheopticaland photoacousticsignals weredonein a similar
mannerasinthecalibrationmeasurement.
Theabsorptioncoefficient
m
aoftheabsorbingbloodsampleinsidethetubemountedintheintegratingsphere1wasderived
usingsimpleenergybalancewithinthesphere.Theincidentlight
energyEinwasdistributedoverandwasabsorbedbythevarious
parts within the sphere, such that Ew, Ea and Eout are the
magnitudesoftheabsorbedenergybythespherewall,absorber
tubeandoutputport,respectively.Fromsimpleenergy balance
considerations,anequationcouldbewrittenasfollows
Ein¼EwþEaþEout (1)
Eq.(1)couldalsobewrittenintermsoftheuniformfluence
F
withinthesphere
Ein¼cw
f
þcaf
þcof
(2)cw,caandcodependonthematerialsused.Forthecaseofaweakly
absorbing sample inside the tube (relatively low absorption
coefficient which is less than 2mm1), c
a=
m
a V, such thatabsorptionwasuniformovertheentirephysicalvolumeV.Using
Eout=co
F
,Eq.(2)becomesEin¼cw Eout co þ
m
a VEout co þ Eout (3) Eq.(3)simplifiesto Ein Eout ¼aþbm
a (4) wherea¼ðcw=coÞþ1andb=V/coMoreover, Ein could bedetermined using Eout if there is no
absorbingsampleinsidethetubesinceEin=cinEout,noabsorber.Also,
Eout¼coutE0out where cout is a constant that takes into account
experimentalfactorssuchassensitivityofdetectionand
conver-sionfromabsolutevaluetoarbitraryunits,forexampleJoulesto
Counts,as wellas theattenuation due tofiber coupling.Thus,
Eq.(4)couldbewrittenas
E0in E0 out ¼AþB
m
a (5) whereE0 in¼E 0out;noabsorber,themeasuredoutputsignalwhenthere
isnoabsorberinsidethetube,andAandBareconstantswhich
couldbemeasuredexperimentallyviaacalibrationprocedureas
describedabove.RearrangingEq.(5)gives
m
a¼ ðE0in=E 0 outÞA B (6) 2.4.G
measurementThecorrespondingGru¨neisenparameter
G
wasmeasuredusingthefollowingequation[8]
G
¼Vppðcþm
aVÞ E0in;PAkm
a(7)
Here, Vpp was the voltage-peak-to-peak amplitude of the
detectedphotoacousticsignalgeneratedbytheabsorbingblood
samplewith
m
aandvolumeV.m
awasdeterminedviasimulta-neousspectrophotometryusingintegratingsphere1asdescribed
above,whereasV=0.0134cm3,thephysicalvolumeofthetube.c
andkaretheinstrumentconstantsdeterminedviathecalibration
methoddescribedinreference[8].E0in;PAistherelativeenergyof
the incident pulse measured by the photodetector. A detailed
derivationofEq.(7)isgiveninRef.[8].
2.5. Calibrationmeasurement
Acalibrationprocedurewasdonetodeterminetheinstrument
constants. Aqueous ink dilutions were used as absorbers with
known
m
a values as measured by the standard transmissionspectrophotometry technique(Shimadzu UV-VIS). BlackEcoline
ink(RoyalTalensEcoline7008265)wasdilutedwithdeionized
anddemineralizedwaterinordertomakeatleastfive
concentra-tionswith
m
avaluesrangingfromabout0.2to2mm1.Eachinkdilutionwasinjectedintothetubeuntilitflowedouttheotherend
toensurethatthesameabsorbingsamplewasmountedinboth
integrating spheres. Simultaneous measurements ofthe optical
outputandphotoacousticsignalsweredonebysynchronizingdata
collection via computer interface using AvaSpec and LabVIEW
software. The AvaSpec software recorded the amount of light
reachingtheoutputportofthespherewhichwasfiber-coupledto
the spectrometer that scans a spectrum between 400nm and
900nm wavelengths. With the assumption that only the
introductionoftheabsorbinginkwaschangedintheintegrating
spheresystem,therelativeinputlightenergyE0in;abswastakenas
the detected output signalwith only water (no absorbingink)
insidethetubewhereastherelativeoutputlightenergyE0
out;abswas
equivalenttothesignalwiththeabsorbinginkinsidethetube.
Absorptionotherthanthatduetoinkabsorptionwasassumedto
bethesameinallmeasurementsandcanbecancelledoutinthe
calculation for
m
a of the absorber in the tube. The necessaryconstantsformeasuring
m
aofsampleswereobtainedfromaplotofE0
in;abs=E0out;abs ratio versus
m
a values. On the other hand, theLabVIEW program recorded the temporal photodetector and
transducer photoacoustic signals. The area under the curve of
thephotodetectorsignalwastakenastherelativeinputenergy
E0
in;PA,whereasthevoltage-peak-to-peakVppofthephotoacoustic
signal was proportional to the initially generated pressure
amplitude. To obtain the necessary instrument constants for
determining
G
,themeasuredVpp=E0in;PAratiowasplottedagainstcorresponding
m
avaluesofinkdilutions.Themeasuredm
avaluesoftheaqueousinkabsorberswereusedinE0
in=E0outversus
m
aandVpp=E0in;PAversus
m
acalibrationplotsasshowninFigs.2and3.3. Results
Examplesofdetectedphotoacousticsignalswithinkabsorbers
andhumanbloodsamplesareshowninFig.2.Ascanbeseenin
Fig.2a,theVppamplitudeofthesignalincreaseswith
m
aoftheaqueous ink calibration samples. The corresponding Vpp=E0in;PA
versus
m
aplotisgiveninFig.3b.Fromthisplot,thecalibrationconstants are k=1.17m3s1 and c=3.55
105m2.
Simulta-neouswithdetectingphotoacousticsignals,thespectrumofthe
opticaloutputsignalfromthesecondsphereisalsomeasuredand
thecorrespondingE0
in=E0outratioforincreasing
m
aisgiveninFig.3a.A linear fit on this plot gives A=1.02 and B=0.21mm1.
Immediately afterobtainingthe calibrationdata, photoacoustic
measurementswithblood samplesinsidethetubearedone for
severaltimes.Exampleofthedetectedsignalsfromwholeblood
samplesisgiveninFig.2b.Eachplotcorrespondstotheaverageof
fivemeasurementsaveragedfrom128pulses.Asareference,the
detectedsignalwithwaterinsidethetubeisalsogiveninFig.2b
The corresponding E0in=E
0
calibrationconstantsAandBareusedinEq.(7)todeterminethe
m
a,bloodofthebloodsample.Thismeasuredm
a,blood,togetherwiththeconstantskandc,V=0.0134cm3andcorrespondingV
pp=E0in;PA
aresubstitutedintoEq.(1)tocompute
G
blood.3.1. Measurementswithfreshbloodsamplesfromthesamedonor
withcorrespondingsetupcalibrationonvariousdays
Absorptionandphotoacousticmeasurementswithfreshblood
samplesfromthesamedonoraredoneatvariousdays,usinganew
calibrationofthesetupeachday.Simultaneouswithabsorption
andphotoacousticmeasurements,oxygensaturation(SO2)values
of each blood sample are measured using anoximeter
(Avoxi-meter).ThevaluesgiveninTable1indicatethatmeasurablevalues
of
m
a,bloodoffreshbloodsamplesfromthesamedonorcanchangeatvariousdayswhich mainlydependontheoxygen saturation
(S02) levels. For the highest measured S02=93%,
m
a,blood=0.4310.009mm1, whereas for the S02=46%,m
a,blood=0.8070.010mm1,whichisconsistentwiththegenerallyreportedvalueat750nm[9].Ontheotherhand,themeasured
G
bloodranges from 0.16 to 0.18 with an average value of
G
blood=0.1660.008. The standard deviation is only 5% of theaveragevaluewhichindicatesthatthemeasureable
G
valueforbloodsamplesfromthesamedonorisrepeatable.Moreover,themeasured
valueof0.166forthishumanbloodsampleisonly4%differentfrom
thereportedvalueofbovineblood[5].
3.2. Measurementswithfreshbloodsamplesfromdifferentdonors
withonesetupcalibration
Table2showsasummaryofthemeasured
m
a,bloodandG
bloodfrom three different donors (labeled A, B, C) using one set of
calibrationplots.Threemeasurementswiththesamewholeblood
samplefromthesamedonoraredonewhichgivesthestandard
deviationvalues. The corresponding hematocrit foreach whole
bloodsampleismeasuredandisalsogiveninTable2.Measured
SO2valuerangesfrom46%to84%.ThelowSO2levelindicatesthat
thebloodsampleisindeedvacuum-sealedandminimallyexposed
toairfromthetimeitisdrawnfromthehealthyhumandonoruntil
it is placed intothe oximeter.Moreover, themeasured
m
a,bloodvaries withthe measured SO2 values. In particular,as the SO2
increases,themeasured
m
a,blooddecreaseswhichisconsistentwiththereportedbehaviorof
m
aat750nmwavelength[9].Infurtherinvestigatingthedependenceof
G
bloodonhematocrit,Fig.4showsaplotofthemeasured
G
bloodversushematocrit.Forzerohematocrit,measured
G
hct=0=0.1200.002whichiswithin10%differentfromthereportedvalueforbovineserum[5].Asshown
inFig.4,themeasuredvalueof
G
increasedto0.1960.010for100%hematocrit,orthecasewhenthereareonlyerythrocytesinsidethe
tube.The generaltrendisincreasing
G
forincreasinghematocrit,except for between47 and60 which showsa slightdecrease in
G
.Moreover,measurementsaredonewithonlytheredbloodcells(RBC)ofthesamplesdrawnfromthesamedonorsA,BandC.Table3
showsthemeasuredvaluesof
m
a,RBCandG
RBCwhicharerelativelyhigherthanthosegiveninTable2.Therelativelyhighervaluesof
m
a,RBCcanbeattributedtotheabsenceofplasmawhichismostly(about 90%) water and has relatively low absorption at 750nm
wavelength whereas the higher
G
RBC values can be due to theFig.3.Exampleofcalibration plotsfor determiningtheconstants (a)A=1.02and B=0.21mm1 (R2=0.99)usedforcalculatingm
aand(b)k=1.17m3s1 and
c=3.55105m2
(R2
=0.98)formeasuringG.
Fig.2.Exampleofdetectedphotoacousticsignalswith(a)aqueousinkdilutionsofknownmaand(b)humanwholebloodsampleinsidethetube(measuredthreetimes).
Measurementwithwater(blackline)isshownforreference.
Table1
MeasuredvaluesofabsorptioncoefficientmaandGru¨neisenparameterGforeachof
the bloodsamples from the same donor. Theabsorption and photoacoustic measurementsaredoneondifferentdaysandcorrespondingly withdifferent calibration ofthe setup.The average G valuefrom thesethree independent measurementsisGblood=0.1660.008.
Measurement ma,mm1(meanSD) G(meanSD)
1 0.8110.093 0.1690.008
2 0.4310.009 0.1570.003
modifiedthermophysicalpropertiesofanensembleofRBCcompared
tothatofwholeblood.Forexample,thethermalexpansioncoefficient
(
b
) ofRBCisrelativelyhigherthan thatofwholeblood,whichisdirectlyproportionalto
G
.3.3. Measurementswithfreshbloodsamplesfromvariousdonors
withcorrespondingsetupcalibrationonvariousdays
Theresultsofthemeasurementsfromvariousdonorsdoneon
variousdaysaresummarizedinTable4.Therangeofvaluesof
measured
G
forfreshhumanbloodsamplesfromfivedonorsvariesfrom0.141to0.177, witha mean valueof 0.162and standard
deviation(SD)of0.016whichis10%ofthemean.Itshouldalsobe
notedthatthelowestmeasuredvalueiswithin20%differentfrom
thehighestmeasuredvalue,whichcanindicateaperson-to-person
variabilityin
G
thatmaybeattributedtovariationsinhematocritandbloodcompositionandcondition.
3.4. Comparisonbetweenmeasurementsatroomandatbody
temperatures
Table5shows
m
aandG
valuesofbloodsamplesmeasuredatroom(228C)andbody(378C)temperatures.Opticalabsorptionof
thesamplesdoesnotchangewithtemperature.Ontheotherhand,
photoacousticefficiency
G
increasedwithtemperature.Forthissetofmeasurements,
G
increasedbyabout70%,from0.1500.011at228Cto0.2450.006at378C.Measurementsat378Caredonefive
times. Table 6 shows a summary of measured values for blood
samplesfromfive differentdonors.Asnotedabove, thevaluesof
measured
m
achangedueto theinherentoxygenationlevelofthebloodsamples.Ontheotherhand,theaverage
G
=0.2260.015at378Chas7%standarddeviationwhichindicatesthatthemeasured
G
atbodytemperaturealsovariesminimallyrelativetomeanvalue,
althoughthelowestmeasuredvalueof0.208isapproximately18%
differentfromthehighestmeasuredvalueof0.245whichissimilarto
thedifferenceobservedat228C.Comparingtheaverage
G
measuredat378Cwiththeonemeasuredat228C,measured
G
increasedbyabout 40% with increased temperature. The increased value of
measured
G
canbemainlyattributedtothecombinedincreaseintherelevantthermalexpansionpropertiesofthedifferentcomponentsof
blood,suchasred bloodcellsandplasmawhichis mostlywater
whichincreaseswithtemperature[3,10].Furthermore,itisrecently
reported that the photoacoustic signal from blood samples with
varying hematocrit increases with temperature which gives an
indicationonthetemperaturedependenceof
G
[11].4. Discussion
Prior to performing measurementswith blood samples,the
influenceoflightscatteringtomeasureablevaluesofabsorption
coefficients and Gru¨neisen parameterof aqueousink solutions
with intralipid was investigated as presented in our paper
[12].Resultsshowthattherangeofvaluesthatcanbeaccurately
measuredusingourintegratingspheremethodare
m
a<1.5mm1and
m
0s<3 mm1,whicharewithintherangeofreportedvalues
formostbiologicalfluidsincludingblood,atinfraredwavelengths.
Photoacoustic measurementswiththeseabsorbingaqueousink
samples (with and without the intralipid) give values of the
Gru¨neisenparameterclosetothatofwater,asexpectedsincethe
Table2
Measuredvaluesofoxygensaturation(SO2),hematocrit(hct),absorptioncoefficientmaandGru¨neisenparameterGforeachofthebloodsamplesfromdonorsA,BandC.
Samples SO2,%(meanSD) Hematocrit,%(meanSD) ma,mm1(meanSD) G(meanSD)
A 634 492 0.6940.026 0.1770.005
B 844 461 0.5800.016 0.1660.006
C 463 471 0.8070.010 0.1730.002
Fig.4.MeasuredGru¨neisenparameterGversusbloodhematocrit.
Table3
Measuredvalues ofabsorptioncoefficientma andGru¨neisenparameterGfor
sampleswith100%hematocrit(containingonlyredbloodcells)fromeachofthe blood samples from donorsA, B and C. Three simultaneous absorption and photoacousticmeasurementsaredoneforeachsampletoobtainthestandard deviation(SD).
Samples ma,mm1(meanSD) G(meanSD)
A 0.9490.022 0.1890.006
B 0.8180.015 0.2070.019
C 0.9060.011 0.1920.009
Table4
MeasuredvaluesoftheGru¨neisenparameter
G for blood samples from five different donors(donorsD1–D5).Threesimultaneous absorptionandphotoacousticmeasurements are done for each sample to obtain the standarddeviation(SD). Samples G(meanSD) D1 0.1770.005 D2 0.1520.020 D3 0.1660.008 D4 0.1760.007 D5 0.1410.003 Table5
maandGvaluesofbloodsamplesfromonedonormeasuredatroom(228C)and
body(378C)temperatures.
Temperature(8C) ma(mm1) G
22 0.5010.028 0.1500.011
37 0.4930.034 0.2450.006
Table6
maandGvaluesofbloodsamplesfromfivedifferentdonorsmeasuredatbody
temperatureof378C. Samplenumber ma(mm1) G 1 0.4930.034 0.2450.006 2 0.7380.003 0.2160.005 3 0.4970.012 0.2080.007 4 0.9080.032 0.2260.005 5 0.9050.012 0.2400.005
samplesarecomposedmostlyofwater.Itshouldbenotedthatthe
absorptioncoefficientcanalsobe measured fromthetemporal
profileofthephotoacousticsignalusingbackwardmodedetection
[13,14].Thisphotoacousticmeasurementoftheopticalabsorption
is also investigated for turbid medium using Monte Carlo
simulations [15]. In this research, it hasbeen shown that the
diameteroftheincidentlaserbeamcanbechosensuchthatthe
absorbed optical energy, which is proportional to the
photo-acoustic amplitude, is linearly dependent on the absorption
coefficient, independent of the scattering coefficient. In our
integratingspheremethod, theincidentlight ishomogeneously
distributed on the absorbing target independent of the beam
diameterfor
m
a<1.5mm1andm
0s<3 mm1.Absorption and photoacoustic measurements were done on
fresh blood samples from healthy human donors. No further
analysesonthesamplesweredonetocheckforsimilaritiesand
differencesintheir physiologicalconditions.Therefore,different
measurement scenarios were designed and implemented to
investigatetheinfluenceofvariationsineitherthebloodsamples
ortheexperimentalsetup.Performingmeasurementsonvarious
days(notnecessarilyatregularintervals)wereaimedatinvestigating
onthemeasurablevaluesof
G
withvariousfreshbloodsamplesusingtheintegratingspheresetupwithnewcalibrationmeasurementfor
everysample.Thiswastoinvestigatethestabilityofthemethodand
setupandhowanyperturbationonthesystem,suchaschangingthe
absorbingsampleordoingnewinstrumentcalibration,couldaffect
the measurable values. Results indicated that the measured
m
achanges with blood oxygenation and the measured
G
of bloodsamplesfromaparticulardonorwasthesameforthreeindependent
measurementswitharelativeerrorofonly5%.
To investigatefurther,measurementswithmorefreshblood
samplesfromvarioushumandonorsaspiratedondifferentdays
wereperformed.Resultsindicatedthatthevaluesofmeasured
G
varyamongfivedifferentdonorswithamaximumdifferenceofup
to20%,whichistwicethemeasuredexperimentalerrorobtained
withcalibrationofthesystem,asdescribedin[8].Thisobservation
suggestedthatthe
G
of humanblood mayvary fromperson topersonwhichdependsonphysiologicalfactors.Unfortunately,no
furtheranalysisonthecompositionandconditionoftheobtained
bloodsampleswasdone.Onepossiblereasonforthedifference
couldbethevariationsinthehematocritofthesamples.However,
based on the results of hematocrit dependence measurements
showninFig.4,the
G
valuesof0.14and0.17wouldcorrespondtohematocrits of 20 and 60, respectively, which are beyond the
normalrangeforhealthywholeblood.Anotherpossiblesourcefor
thevariationinmeasured
G
couldbethevariationinthesampletemperature.Fromthemeasuredaveragevalues,anincreasefrom
0.16at228Cto0.22at378Cimpliesthat0.004increasein
G
perdegree increase in temperature. This would give a temperature
fluctuationof10degreesbetween
G
=0.14and0.17,whichisveryunlikelysincethemeasuredambienttemperatureduringtheentire
experimentvarieswithin onedegreeonly.Therefore,variationsin
hematocritandtemperature,althoughtheymayhaveasmalleffect,
werenotthemainreasonfortheobservedvariationin
G
ofbloodfromvarioushuman donors.Moremeasurementscan bedoneto
exploreonthevariabilityof
G
fromperson-to-personandtofurtherinvestigatetherelevantphysiologicalfactorsaffectingthis.
ResultsshowninFig.4indicatedthatthemeasuredvalueof
G
generallyincreaseswithbloodhematocrit.Bloodsamplewithzero
hematocritcorrespondstosamplewhich containsbloodplasma
and that with 100% hematocrit corresponds to sample which
containsonlyredbloodcells.Thedifferenceinthevaluesof
G
forthese two samples (with zero and 100% hematocrit)could be
mainly attributed to the difference in the composition (with
different thermophysical properties). For the samples with
hematocritbetween zeroand100%, thecombinationof plasma
andredbloodcellsmayhavedifferentthermophysicalproperties
whichgiveadifferenteffective
G
values,moreredbloodcellsmayindicatehigher
G
.Ithasalsobeenreported(asdescribedinRef.[16]) that scattering effects could result to higher value of
calculated
G
.Theobservedvariationin
G
withhematocritandtemperaturecould have consequences for in vivo photoacoustic imaging of
microcirculationofblood.Smallerbloodvesselstendtohavelower
hematocrit, the so-called Fahraeus effect, which lowers the
effectivebloodviscosityinthesmallervesselsofthe
microcircu-lation,especiallyinneonates[17,18].
The dependenceof
G
on blood hematocrit and temperaturecouldformacomplicationforquantitativephotoacousticimaging,
with the aim to determine true absorption coefficients. For
example,basedon theobtained results,a changein hematocrit
values from50% to20%, whichisfoundinthemicrocirculation
whencomparingbloodinlargemicrocirculatoryvesselswithsmall
vessels,coulddecreasetheGru¨neisenparameterbyapproximately
20%.Furthermore,afive-degreevariationoftemperaturewithin
thebody couldinduceatleast 10%variations oftheGru¨neisen
parameter.Such afive-degreetemperaturedifference mayexist
betweenthebodycoreandtheskin,forinstance.Hence,natural
variations in local temperature or hematocrit may cause a
significant variation in the Gru¨neisen parameter. Since the
photoacousticstress
s
oisrelatedtoG
,m
aandopticalfluenceF
as
s
o=Gm
aF
, thesenaturalvariationsin thevalueofG
couldintroduceanextrauncertaintyintotheproblemofquantifying
m
a,ontopoftheuncertaintyinfluence
F
.Itshouldalsobenotedthatinourmeasurementswithaqueous
ink and blood samples, the difference in acoustic impedance
betweencalibrationmediumandbloodmediumwasnotexplicitly
considered in the calculations. The polyethylene tube used to
mountthesampleinsidetheintegratingspheresystemwaschosen
becauseithasacousticimpedancesimilartothatofsoftbiological
tissues.Eventhoughtheacousticimpedanceisnotconsideredin
the calculations, the
G
values we obtained are in quite goodagreementwiththatforbovinebloodreportedinrelatedliterature.
Moreover,theobservationsforvaryinghematocritandtemperature
are notaffected by a potential difference of acoustic impedance
betweenbloodandthecalibrationmedium.Alsothevariationsof
G
betweensubjects,orbetweendays,arenotaffectedbythis.
5. Conclusion
Amethodandsystemfordeterminingtheabsorption
coeffi-cient
m
a and Gru¨neisen parameterG
liquid absorbing andscatteringsamplesweredesignedandimplementedusingcoupled
integratingspheres.Onespherewasusedasaplatformfordoing
absorptionmeasurementsandanothersphereforphotoacoustic
measurementswiththesampleinside atubemounted
simulta-neouslythroughbothspheres.Usingthemeasuredrelativeoptical
outputratios,alinearequationfordetermining
m
awasderivedbasedonsimpleenergybalancewithinthesphere.Thismeasured
m
awasusedincalculatingforG
ofthesample.Theapplicationofthe developed platform referred as quantitative photoacoustic
integrating sphere (QPAIS)wasdemonstrated bymeasuring
m
aand
G
ofhumanblood.Atroomtemperaturemeasurements,foraparticular donor, the measured
m
a of blood ranged from0.807mm1 to 0.431mm1 with S0
2 values of 46% and 93%,
respectivelyand corresponding measured
G
=0.1660.008 wasrepeatable for three independent measurements withnew setup
calibrationatvariousdays.Thisvalueisingoodagreementwiththe
valuesobtainedforbovineblood[5].Moremeasurementswithfresh
bloodsamplesfromvariousdonorsindicatedadecreasing
m
avalueforincreasingbloodoxygenationlevelswhichisconsistentwiththat
blood samples fromfive human donors, withapproximately 20%
differencebetweenlowestvalue0.141andhighestvalue0.177,each
within15%differencefromthemean.Measurementswithvarying
bloodhematocritindicatedthat
G
increaseswithbloodhematocrit.However,thevariationinwholebloodhematocritwastoosmallto
causetheobserved20%variationin
G
.Thebloodsamplesobtainedwereassumedtobefromhealthydonors.Theactualphysiological
stateofthebloodsamplesanditseffectonthemeasurable
G
couldbefurtherinvestigated.Moreover,measurementsatbodytemperature
of 378C gave an average
G
=0.2260.015 which is about 40%different from the measured value at 228C. This observation of
increasing
G
valuewithincreasingtemperaturewasconsistentwithresultsinRef.[11].
The shown dependence of Gru¨neisen parameter on blood
hematocritandtemperaturecouldformanextracomplicationfor
quantitativephotoacoustics,becauseofthenatural variationsof
hematocrit found in the microcirculation, and temperature
differencesbetweenthebodycoreandtheskin.
Themethodpresentedherecouldbeusedformeasuringwith
other weakly absorbing liquid samples which are relevant to
biomedicine,particularly the target absorbers in photoacoustic
imaging.Itshouldbenotedthatthedemonstrationofthemethod
waspresentedhereusingonlyonewavelength.Inprinciple,QPAIS
could be used with a range of wavelengths such that further
quantitative investigations could be done. Additionally, the
requiredincidentenergy perpulsewasrelativelylow suchthat
thelasersourcecouldbechangedtoaportablelaserdiodewith
reasonablyshortpulseduration.Alightsourcewithlessenergyper
pulse(approximately<3mJperpulse)couldbeused,insteadof
thehighenergysourcesusedforphotoacousticimaging.
Conflictofinterest
Nonedeclared.
Acknowledgements
TheauthorsacknowledgeTheNetherlandsTechnology
Foun-dationSTWforthefinancialsupporttothisresearch(Vicigrant
10831).Likewise,theauthorssincerelyappreciate thekindand
generous assistance of the Experimental Center for Technical
Medicine(ECTM)oftheUniversityofTwente,togetherwiththe
varioustechniciansanddonors,whofacilitatedthesupplyoffresh
humanbloodsamplesusedintheabsorptionandphotoacoustic
measurements. References
[1]B.Cox,etal.,Quantitativespectroscopicphotoacousticimaging:areview,J. Biomed.Opt.17(6)(2012).
[2]V.E.Gusev,A.Karabutov,LaserOptoacoustics, AmericanInstituteofPhysics, NewYork,1993.
[3]L.V.Wang,H.A.Wu,BiomedicalOptics:PrinciplesandImaging, JohnWiley& SonsInc.,NewJersey,2007.
[4]E.V.Savateeva,etal.,Opticalpropertiesofbloodatvariouslevelsof oxygenationstudiedbytimeresolveddetectionoflaser-inducedpressure profiles,Biomed.Optoacoust.III4618(2002)63–75.
[5]D.K.Yao,etal.,PhotoacousticmeasurementoftheGruneisenparameterof tissue,J.Biomed.Opt.19(1)(2014).
[6]B.Soroushian,W.M.Whelan,M.C.Kolios,Studyoflaser-inducedthermoelastic deformationofnativeandcoagulatedex-vivobovinelivertissuesfor estimatingtheiropticalandthermomechanicalproperties,J.Biomed.Opt.15 (6)(2010).
[7]P.Elterman,Integrating-spherespectroscopy,J.Opt.Soc.Am.59(11)(1969)1537.
[8]Y.Villanueva,etal.,PhotoacousticmeasurementoftheGruneisenparameter usinganintegratingsphere,Rev.Sci.Instrum.85(7)(2014).
[9]N.Bosschaart,etal.,Aliteraturereviewandnoveltheoreticalapproachonthe opticalpropertiesofwholeblood,LasersMed.Sci.29(2)(2014)453–479.
[10]I.Larina,K.Larin,R.Esenaliev,Real-timeoptoacousticmonitoringof temperatureintissues,J.Phys.D:Appl.Phys.38(2005)2633–2639.
[11]E.V.Petrova,A.A.Oraevsky,S.A.Ermilov,Redbloodcellasauniversal optoacousticsensorfornon-invasivetemperaturemonitoring,Appl.Phys.Lett. 105(2014).
[12]Y.Y.Villanueva,C.Veenstra,W.Steenbergen,Measuringabsorptioncoefficient ofscatteringliquidsusingatubeinsideanintegratingsphere,Appl.Opt.55 (15)(2016).
[13]A.Karabutov,N.B.Podymova,V.S.Letokhov,Time-resolvedlaseroptoacoustic tomographyofinhomogeneousmedia,Appl.Phys.B63(1996)545.
[14]A.Karabutov,etal.,Backwardmodedetectionoflaser-inducedwide-band ultrasonictransientswithoptoacoustictransducer,J.Appl.Phys.87(4)(2000).
[15]I.M.Pelivanov,etal.,Opto-acousticmeasurementofthelocallightabsorption coefficientinturbidmedia:1.Monte-Carlosimulationoflaserfluence distributionatthebeamaxisbeneaththesurfaceofaturbidmedium, QuantumElectron.39(9)(2009)830–834.
[16]R.K.Saha,M.C.Kolios,Photoacousticsignalsfromredbloodcells,J.Acoust. Soc.Am.129(5)(2011).
[17]J.H.Barbee,G.R.Cokelet,TheFahraeuseffect,Microvasc.Res.3(1)(1971)6–16.
[18]E.P.Zilow,O.Linderkamp,Viscosityreductionofredbloodcellsfrompreterm andfull-termneonatesandadultsinnarrowtubes(Fahraeus-Lindqvisteffect), Pediatr.Res.25(6)(1989)595–599.
YolandaVillanueva-Paleroobtainedherbachelorof scienceinappliedphysicsdegreefromtheNational InstituteofPhysicsattheUniversityofthePhilippines in 2000. She did her bachelor thesis project on holographicopticaldataencryptioninaniron-doped lithiumniobatecrystalatthePhotonicsResearchgroup. InDecember2003,shealsojoinedasix-monthresearch trainingprogramattheformerPrecisionInstrument Development Center of Taiwan Republic of China, wheresheperformedpulsedlaserdepositionofzinc oxide thinfilms. She obtained a masterof physics degreefromtheVrijeUniversityAmsterdaminAugust 2010.HermasterthesiswasonCasimir-likeeffectin granularfluids.InJanuary2016,sheobtainedherPhD degreeattheUniversityofTwentewhereshedevelopedaquantitativephotoacoustic integratingsphereplatformformeasuringtheGru¨neisenparameter,opticalabsorption andfluorescencequantumyieldofbiomedicalfluids,underthesupervisionofProf. WiendeltSteenbergenofthebiomedicalphotonicimaginggroup.
ErwinHondebrink receivedhis bachelordegreein biomedicalelectrical engineering in 1998 at Hoge-schoolEnschede.From1999to2015,heworkedalmost continuouslyasaresearchengineerattheUniversityof Twente.HeworkedintheLowTemperaturegroupand the Biomedical Imaging group on various research projectsinmicrocooling,laserdoppler,LASCA, photo-acousticsandacoustooptics.Hisfocusisonsoftware development,dataacquisition,visionandelectronics. In2007,heworkedatPerimedABinSwedenonthe developmentofaLASCAsystem.From2009to2010,he workedatOstendumBVonaportablebiosensorforthe detectionofbacteria,viruses,yeasts,andbiomarkers. WilmaPetersenstudiedmedicallaboratoryeducation withthespecializationMedicalChemistryandClinical ChemistryattheHogeschoolEnschede.Shereceived bothdegreesin1998.ShestartedworkinginAcademic Medical Center (AMC) in Amsterdam as research analystwereshegeneratedproteinstodostructural functionanalysis.In2005,shejoinedtheBiomedical PhotonicImaginggroupattheUniversityofTwente, Enschedewhereshepreparedphotoacousticcontrast agentssuchasgoldnanorodsthatwereconjugatedwith antibodies.Variousaspectsofthesecontrast agents wereresearchedandinvestigated.Sheisnowworking onseveralotherprojectsincludingthephotoacoustic measurement ofthe Gru¨neisenparameterusing an integratingsphere.
WiendeltSteenbergenobtainedaPhDdegreeinfluid dynamicsattheEindhovenUniversityofTechnologyin 1995,afterwhichhejoinedtheUniversityofTwente, Enschede(theNetherlands)asapostdoc.In2000he wasappointedassistantprofessorinbiomedicaloptics andbroadenedhisscopetolow-coherence interferom-etryandphotoacousticandacousto-opticimaging.In 2010,hebecamefullprofessorandgroupleaderofthe newlyformedBiomedicalPhotonicImaginggroupof theUniversityofTwente.