Extended in-vitro protocols for TEHV : implanted by minimally
invasive procedures
Citation for published version (APA):
Dijkman, P. E., Driessen - Mol, A., Stenger, R., Baaijens, F. P. T., & Hoerstrup, S. P. (2008). Extended in-vitro protocols for TEHV : implanted by minimally invasive procedures. Poster session presented at Mate Poster Award 2008 : 13th Annual Poster Contest.
Document status and date: Published: 01/01/2008 Document Version:
Publisher’s PDF, also known as Version of Record (includes final page, issue and volume numbers) Please check the document version of this publication:
• A submitted manuscript is the version of the article upon submission and before peer-review. There can be important differences between the submitted version and the official published version of record. People interested in the research are advised to contact the author for the final version of the publication, or visit the DOI to the publisher's website.
• The final author version and the galley proof are versions of the publication after peer review.
• The final published version features the final layout of the paper including the volume, issue and page numbers.
Link to publication
General rights
Copyright and moral rights for the publications made accessible in the public portal are retained by the authors and/or other copyright owners and it is a condition of accessing publications that users recognise and abide by the legal requirements associated with these rights. • Users may download and print one copy of any publication from the public portal for the purpose of private study or research. • You may not further distribute the material or use it for any profit-making activity or commercial gain
• You may freely distribute the URL identifying the publication in the public portal.
If the publication is distributed under the terms of Article 25fa of the Dutch Copyright Act, indicated by the “Taverne” license above, please follow below link for the End User Agreement:
www.tue.nl/taverne
Take down policy
If you believe that this document breaches copyright please contact us at:
openaccess@tue.nl
providing details and we will investigate your claim.
Extended in-vitro protocols for TEHV
-Implanted by Minimally Invasive
Procedures-Petra Dijkman1, Anita Mol1, Rene Stenger2, Frank Baaijens1, Simon P. Hoerstrup1,2
1 Eindhoven University of Technology, Department of Biomedical Engineering, The Netherlands;
2 University Hospital Zürich, Clinic for Cardiovascular Surgery and Department of Surgical Research, Switzerland
/ Biomechanics & Tissue Engineering
Introduction
Tissue engineering of heart valves, based on rapid degrading polymer scaffolds, and minimally invasive valve replacement procedures represent promising technologies for patients with valvular heart disease. The successful merging of these novel technologies was demonstrated in a large animal model previously [1]. However, local irregularities and thickening of the TEHV were observed after explantation (Fig. 4), of which the first may be partly due to separation of the leaflets shortly before implantation (Fig.1&2).
We hypothesized that overgrowth of the separation areas before implantation could reduce the in-vivo development of irregularities on the leaflets tips. Therefore, we investigated the influence of an advanced in-vitro methodology on valve performance.
Moreover, improved in-vivo performance (initial pressure gradients were 10 mmHg instead of 20 mmHg) was evident directly after implantation. However, the explant outcome was similar to the previous explants (Fig.4) and excessive remodeling was observed.
Figure 2: Separation of the leaflets, crimping of the TEHV, and the TEHV inserted into the implantation device.
Figure 3: Typical results of H&E (A) and Masson trichrome (B) staining of control leaflets of the advanced culture protocol. Separation areas in the leaflets are overgrown by tissue. Scale bars represent 500 µm.
Discussion
The improved in-vitro methodology, to reduce the in-vivo thickening response, covered the leaflet tips with a sufficient tissue layer. Moreover, the initial in-vivo valve performance was promising. Nevertheless, there was no improvement regarding the excessive remodeling previously found after explantation.
Although representing only a preliminary in-vivo study, we conclude that the local irregularities after implantation are not directly related to the separation of the leaflets and further research is necessary. Moreover, extended studies have to be initiated to clarify the in-vivo remodeling response. References
[1] Dijkman PE et al. AHA scientific sessions 2007 [2] Mol A et al., Ann Biomed Eng 2005; 33:1778-88
Figure 4: Typical results of H&E (A), Masson trichrome (MT; B), and α-SMA (C) staining of explanted TEHV. H&E staining demonstrates cellular tissue formation (A). MT reveals abundant amounts of collagen (blue, B). α-SMA produced by active fibroblasts is shown in brown (C). Arrows indicate irregularities, stars indicate the implanted tissue. Explants do not reveal large differences independent of the in-vitro methodology. Scale bars represent 200 µm.
A
B
A
B
Results
The advanced culture methodology indeed demonstrated the expected overgrowing of the separation areas (Fig.3).
Methods
Trileaflet heart valves (n=4, ØID28mm), based on rapidly degrading polymer scaffolds and self-expandable stents, were engineered from ovine vascular derived autologous cells. Valves were grown in-vitro for 13 days, as described previously [1]. To overgrow the separation areas, TEHV were cultured for an additional 6 days after separation of the leaflets utilizing combined strain-flow bioreactor systems (an adaptive version of the diastolic pulse duplicator systems [2]). After crimping (Fig.1) two valves were delivered minimally invasively (mini-thoracotomy, trans-apical approach) in pulmonary position in sheep. Valves were explanted after 4 and 7 weeks.
Photograph: Bart van Overbeeke
Figure 1: Typical results of H&E (A) and Masson trichrome (B) staining of control leaflets of the previous culture protocol. Separation areas in the leaflets are indicated by arrows. Scale bars represent 500 µm.