116

Abstract no. 97

Vascular Biology

Continuous vital signs monitoring using the VitalConnect MD Healthpatch

M.S. van Mourik[1], M.C. Hermans[1,2], H.J. Hermens[2], J. Baan Jr.[1], M.M. Vis[1]

[1]: AMC Heart Center, University of Amsterdam

[2]: MIRA Institute of Technical Medicine and Biomedical Engineering, University of Twente

Background

The field of mobile sensor technology is rapidly developing and tools enabling simultaneous monitoring of multiple physiological parameters are emerging. Hence, wearable sensors might be used for quantification of general cardiac condition or for purposes of telemonitoring. To explore the usability of VitalConnect MD Healthpatch (HP) as a potential tool for remote patient monitoring, we verified signal data quality of monolead ECG, heart rate (HR) and respiratory frequency (RF). Furthermore first experience and signal acquisition during monitoring was performed in monitoring in patients .

Methods and results

Five healthy subjects were monitored by the HP during various activities including rest, metronome breathing, cycling and daily life activities. The ECG morphology, HR and RF were validated using 6-lead reference ECG and respiratory flow. Visibility of P-waves and artifacts was scored by 3 investigators independently. Additionally, HP recordings were made in 3 patients undergoing transfemoral TAVI peri-procedurally. The ECG showed good correspondence with the reference ECG during most tasks and involved a mean maximal correlation of 0.9 in rest. P-waves were visible in 72-100% of the time, but could not be distinguished for all recommended patch locations. Artifacts were seen during arm movement and sensor disturbances. The HR and RF parameters were associated with a mean Pearson’s coefficient of 0.94 and 0.64 respectively, and were able to represent the trends in the data with a minimal delay. ECG abnormalities appearing in the patients, including atrial fibrillation, (premature) ventricular complexes and AV- or bundle branch blocks, could be identified in the HP recordings.

Conclusion

The HP provides accurate monolead ECG and reflects global HR and RF changes during most activities if placed in specific chest position. The Healthpatch might particularly be suitable for identification of ECG abnormalities and trends in HR and RF, and hence be valuable for purposes of remote monitoring.

Referenties

GERELATEERDE DOCUMENTEN

This study is contributing to a growing body of literature on continuous auditing by answering the following research question: “What are the managerial motives of using

Instead of needing care-providers around all the time, the proposed solution is to create and provide a Home Care System that can monitor many things and that includes an

Copyright and moral rights for the publications made accessible in the public portal are retained by the authors and/or other copyright owners and it is a condition of

"VXVY" verwerkingsprogramma voor statische metingen met een 16-kanaals scanner en een DANA universeelmeter.. Citation for published

Indien weinig acht behoeft te worden geslagen op de letter- structuur van een woord, dan lijkt de kans groot dat tijdens lees- oefeningen waarbij naar wens de gesproken woordklank

The relative estimation errors in cardiorespiratory interactions were evaluated for different types of apneas and respiratory signals, and the lowest errors were obtained for ˆr rs

Het gegeven dat de oppervlakte van de grootste cirkel vijf keer zo groot is als van de kleinste cirkel, betekent dat de straal 5 keer zo groot is. De diagonaal in een

As dialysis patients at haemodialysis are regularly in-hospital (i.e., at least 3 days per week), with a duration of four hours per session, the prediction is only applied for the