• No results found

Sweeping flow electrophoresis (SFE): A new continuous separation technique

N/A
N/A
Protected

Academic year: 2021

Share "Sweeping flow electrophoresis (SFE): A new continuous separation technique"

Copied!
3
0
0

Bezig met laden.... (Bekijk nu de volledige tekst)

Hele tekst

(1)

SWEEPING FLOW ELECTROPHORESIS (SFE): A

NEW CONTINUOUS SEPARATION TECHNIQUE

P. Vulto1, D. Kohlheyer2, G.A. Urban1 and R.B.M. Schasfoort2

1 Laboratory for Sensors, Department of Microsystems Engineering

(IMTEK), Albert-Ludwigs-Universität Freiburg, GERMANY

2 MESA+ Institute for Nanotechnology, University of Twente, Enschede,

THE NETHERLANDS

ABSTRACT

In this paper, a new principle for continuous electrophoretic separation is intro-duced. An electrokinetically pinched sample flow is swept through a microfluidic chamber. The wavelengths of the resulting sine-waves are in direct relation with the electrophoretic mobility of the compound. The principle is applied to separate Rho-damine B and Fluorescein.

KEYWORDS: Sweeping flow electrophoresis, SFE, EOF INTRODUCTION

Free-flow electrophoresis has several advantages over capillary electrophoresis: the mobility of a separated compound can be determined time-continuously, plug-formation is not required and the analytes can be recovered after separation. A dis-advantage of free-flow electrophoresis is the fact that two different actuation forces are required. These are typically pressure driven flow in forward direction and elec-trokinetic separation in lateral direction [1]. Here we introduce a continuous flow electrokinetic separation principle that is fully electrically controlled. The principle uses a pinched sample flow that is swept through a chamber by two guiding streams. The electrophoretic mobilities of the sample compounds are in direct relation with the wavelength of the resulting sinusoidal separation profiles. We call this technique sweeping flow electrophoresis (SFE).

THEORY

Recently, we introduced an EOF driven technique to position a sample flow in a laminar flow chamber [2]. The principle is depicted in figure 1. Two guiding flows position a central sample stream in a laminar flow chamber. The three flow streams are electroosmotically manipulated, so that control of the applied potentials results in the accurate positioning of the sample stream [2]. The guiding stream potentials are chosen such, that a sinusoidal sample flow is generated. The y-position of the sample stream depends on the time (t), the x-position and the sweeping frequency (f):

( )

⎟ ⎠ ⎞ ⎜ ⎝ ⎛ = λ π π ft x Y t x y , sin 2 2 (1) 978-0-9798064-1-4/µTAS2008/$20©2008CBMS 823

Twelfth International Conference on Miniaturized Systems for Chemistry and Life Sciences October 12 - 16, 2008, San Diego, California, USA

(2)

and

(

)

f E elec eof μ μ λ= + (2)

with Y the amplitude of the sine wave, λ the wavelength of the sine wave, µeof the

electroosmotic mobility of the liquid, µelec the electrophoretic mobility of the sample

compound and E the electric field. In [2] we demonstrated that it is feasible to sweep the sample stream, while keeping the electric field in the chamber constant. For this case λ becomes a constant that depends directly on the electrophoretic mobility of the separated compounds. A periodic image capture in phase with the sweeping fre-quency allows measurement of the wavelength.

Figure 1: Principle of sample stream guiding in a laminar flow chamber. Two guid-ing streams sweep a central sample stream through the chamber

EXPERIMENTAL

A glass microfluidic chip is used as described in [2] (see figure 2). The chip is wet-etched with HF and contains powderblasted interface holes. Glass-glass direct wafer bonding was performed by applying pressure and annealing at 600°C. The chip is placed in a holder that contains reservoirs and electrodes for electrokinetic actuation. The principle is demonstrated for the separation of Rhodamine B and Fluorescein. The sample stream is swept with a frequency of 0.5 Hz.

a b

Guiding stream reservoirs Sample stream reservoir Laminar flow chamber Outlet

Figure 2 (a): Glass microfluidic chip with powderblasted holes. (b) The chip con-tains three inlet holes (one for the sample and two for the guiding streams) and one

outlet hole. A positive potential is applied via external electrodes to the inlet holes. The outlet reservoir is grounded.

824

Twelfth International Conference on Miniaturized Systems for Chemistry and Life Sciences October 12 - 16, 2008, San Diego, California, USA

(3)

RESULTS AND DISCUSSION

Figure 3 shows images of the separated compounds for various parts of the sweeping period. As can be seen, the two compounds are separated from each other by a clear variation in the wavelength of the sine. Fluorescein shows a much shorter wavelength than Rhodamine B. This is in good agreement with the above formulas, since Fluorescein is typically negatively charged, while Rhodamine B is relatively neutral for the buffer used.

t=1/f t=1/(4f) t=1/(2f) t=3/(4f)

x y

Rhodamine B Fluorescein

Figure 3: Sequence of separation images taken at one, one fourth, one halve and three quarter of the sweeping period.

CONCLUSIONS

A new, electroosmotically steered, continuous flow separation technique has been demonstrated. The technique uses two guiding streams to sweep a sample through a microchamber. Wavelength analysis of resulting sine waves yields infor-mation on the various electrophoretic mobilities. Rhodamine B and Fluorescein have been successfully separated. The technique can be considered an alternative to con-ventional capillary chip electrophoresis and free-flow electrophoresis.

ACKNOWLEDGEMENTS

Authors would like to thank Stefan Schlautmann for fabrication of microfluidic chips.

REFERENCES

[1] D. Kohlheyer, G. A. J. Besselink, S. Schlautmann, R. B. M. Schasfoort, Free-flow zone electrophoresis and isoelectric focusing using a microfabricated glass device with ion permeable membranes, Lab Chip, vol. 6, pp. 374 - 380, 2006.

[2] G. A. J. Besselink, P. Vulto, R. G. H. Lammertink, S. Schlautmann, A. v. d. Berg, W. Olthuis, G. H. M. Engbers, R. B. M. Schasfoort, Electrophoresis, Electroosmotic guiding of sample flows in a laminar flow chamber, vol. 25, pp. 3705-3711, 2004.

825

Twelfth International Conference on Miniaturized Systems for Chemistry and Life Sciences October 12 - 16, 2008, San Diego, California, USA

Referenties

GERELATEERDE DOCUMENTEN

Numerous studies of flame interaction with a single vortex and recent simulations of burning in vortex arrays in open tubes demonstrated the same tendency for the turbulent burning

A first class of extensions is suggested by the many differential equations that can be written in terms of Wasserstein gradient flows, as explained in Section 3: can these also

Naar alle waarschijnlijk zijn deze voorwerpen afkomstig uit de graven, ontdekt door ].. Floren en dit nadat

Waarderend en preventief archeologisch onderzoek op de Axxes-locatie te Merelbeke (prov. Oost-Vlaanderen): een grafheuvel uit de Bronstijd en een nederzetting uit de Romeinse

We now focus on the difference between the N-DANSE and DANSE algorithms and the optimal fusion vectors for the case of noisy links, which could be obtained if nodes would have access

This is a sample plain XeTeX document that uses tex-locale.tex and texosquery to obtain locale infor- mation from the operating system..

In the slide content, the following commands can be used in order to specify the content of the overlays: \visible.. Andreas Nolda | The overlays package

(Angrily.) My document class does not work.. The cat has destroyed